Datasheet
Data Sheet
www.tektronix.com10
Frequency
Characteristic Description
Gate Time 131 ms
Measurement Method AC-coupled input using the AC voltage measurement
function
Settling Considerations When measuring frequency after a DC offset voltage
change, errors may occur. For the most accurate
measurement, wait up to 1 second to allow input-blocking
RC time constant to settle
Measurement
Considerations
To minimize measurement errors, shield inputs from
external noise when measuring low-voltage, low-frequency
signals
Accuracy
Range
Frequency
Uncertainty Temperature Coefficient/°C
Outside 18– 28°C
90 days 1 year
23°C ±5°C 23°C ±5°C
100 mV to 750 V
*4, 5
20 Hz – 2 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
2 kHz – 20 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
20 kHz – 200 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
200 kHz – 1 MHz 0.01 + 0.004 0.01 + 0.006 0.002 + 0.002
*4
Input >100 mV.
*5
Limited to 8 × 10
7
V Hz.
Continuity
Characteristic Description
ContinuityThreshold 20 Ω
Test Current 1 mA
Response Time 100 S/s with audible tone
Rate Fast
Maximum Reading 199.99 Ω
Resolution 0.01 Ω
DiodeTest
Characteristic Description
Response Time 100 S/s with audible tone
Rate Fast
Maximum Reading 1.9999 V
Resolution 0.1 mV