Datasheet

Digital Multimeters Tektronix DMM4020
Resistance
Specications
are for 4-wire resistance function, or 2-wire resistance with REL. If
REL is not used, add 0.2 for 2 -wire resistance plus lead resistance.
Characteristic
Description
Measurement Method
Current source referenced to LO input
Max Lead Resistance
(4-wire ohms)
10% of range per lead for 200 ,2k ranges. 1 k
per lead on all other ranges
Input Protection 1000 V on all ranges
Input Characteristics
ResolutionRange
Full Scale
(5½ Digits)
Slow
Medium Fast
Current Source
200 199.999 0.001 0.01 0.01
0.8 mA
2k 1.99999 k 0.01 0.1 0.1
0.8 mA
20 k 19.9999 k 0.1 1 1
0.08 mA
200 k 199.999 k 1 10 10
0.008 mA
2M 1.99999 M 10 100 100
0.9 μA
20 M 19.9999 M 100 1k 1k
0.16 μA
100 M 100.000 M 1k 10 k 10 k 0.16 μA||10M
Accuracy
Uncertai
nty*
3
90 days 1 year
Range
23 °C ±5 °
C
23 °C ±5 °
C
Temp erat
ure Coefcient/°C
Outside 1
8-2C
200
0.02 + 0.004 0.03 + 0.004 0.003 + 0.0006
2k
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
20 k
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
200 k
0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
2M
0.03 + 0.003 0.04 + 0.004 0.004 + 0.0005
20 M
0.2 + 0.003 0.25 + 0.003 0.01 + 0.0005
100 M
1.5 + 0.004 1.75 + 0.004 0.2 + 0.0005
*
3
Uncertainty given as ±(% of reading + % of range).
www.tektronix.com/dmm 7