Datasheet
Data Sheet
AC Voltage Specifications
AC Voltage spec
ifications are for AC sinewave signals >5% of range. For inputs from
1% to 5% of range and <50 kHz, add an additional error of 0.1% of range, and for
50 kHz to 100 kHz, add 0.13% of range.
Characteristic
Description
Maximum Input 750 V
RMS
or 1000 V peak or 8 × 10
7
Volts-Hertz product
Measurement Method
AC-coupled true RMS. Measures the AC component of
input with up to 1000 VDC bias on any range
AC Filter Bandwidth
20 Hz - 100 kHz
Common Mode
Rejection
60 dB at 50 Hz or 60 Hz (1 kΩ unbalance)
Maximum C rest Factor 3:1 at Full Scale
Additional Crest Factor
Errors (<100 Hz)
Crest Factor 1-2, 0.05% of full scale
CrestFactor2-3,0.2%offullscale
Only applies for non-sinusoid signals
Input Characteristics
ResolutionRange
Full Sca
le
(5½ Digi
ts)
Slow
Medium Fast
Input Impe dan ce
200 mV 199.999 mV 1 μV10μV10μV
2 V 1.99999 V 10 μV100μV100μV
20 V 19.9999 V 100 μV1000μV 1000 μV
200 V 199.999 V 1 mV 10 mV 10 mV
750 V 750.00 V 10 mV 100 mV 100 mV
1MΩ ±2% shunted by
<100 pf
Accuracy
Uncertainty*
3
90 d ay
s
1 year
Range Frequ
ency
23 °C ± 5 °C 23 °C ±5 °C
Temperature C oef ficien t/°C
Outside 18 - 28 °C
20 Hz -
45 Hz
0.8 + 0
.05
0.9 + 0
.05
0.01 +
0.005
45 Hz -
20 kHz
0.15 +
0.05
0.2 + 0
.05
0.01 +
0.005
20 kH
z-50kHz
0.3 +
0.05
0.35
+0.05
0.01
+ 0.005
200 mV
50 kH
z-100kHz
0.8 +
0.05
0.9 +
0.05
0.05
+0.01
20 Hz
-45Hz
0.8 +
0.05
0.9 +
0.05
0.01
+ 0.005
45 Hz
-20kHz
0.15
+0.05
0.2 +
0.05
0.01
+ 0.005
20 kH
z-50kHz
0.3 +
0.05
0.35
+0.05
0.01
+ 0.005
2V
50 k
Hz - 100 kHz
0.8
+0.05
0.9
+0.05
0.0
5+0.01
20 H
z-45Hz
0.8
+0.05
0.9
+0.05
0.0
1 + 0.005
45 H
z-20kHz
0.1
5 + 0.05
0.2
+0.05
0.0
1 + 0.005
20 k
Hz - 50 kHz
0.3
+0.05
0.3
5 + 0.05
0.0
1 + 0.005
20 V
50 k
Hz - 100 kHz
0.8
+0.05
0.9
+0.05
0.0
5+0.01
20 H
z-45Hz
0.8
+0.05
0.9
+0.05
0.0
1 + 0.005
45
Hz - 20 kHz
0.
15 + 0.05
0.
2 + 0.05
0.
01 + 0.005
20
kHz - 50 kHz
0.
3 + 0.05
0.
35 + 0.05
0.
01 + 0.005
200
V
50
kHz - 100 kHz
0.
8 + 0.05
0.
9 + 0.05
0.
05 + 0.01
20
Hz - 45 Hz
0.
8 + 0.05
0.
9 + 0.05
0.
01 + 0.005
45
Hz - 20 kHz
0.
15 + 0.05
0.
2 + 0.05
0.
01 + 0.005
2
0kHz-50kHz
0
.3 + 0.05
0
.35 + 0.05
0
.01 + 0.005
75
0V
5
0kHz-100kHz
0
.8 + 0.05
0
.9 + 0.05
0
.05 + 0.01
*
3
Uncertainty given as ±(% of reading + % of range).
6 www.tektronix.com/dmm