Datasheet
Data Sheet
Frequency
Characteristic Description
Gate Time
131 ms
Measurement
Method
AC-coupled input using the AC voltage
measurement function
Settling
Considerations
When measuring frequency after a DC offset
voltage change, errors may occur. For the most
accurate measurement, wait up to 1 second to allow
input-blocking RC time constant to settle
Measurement
Considerations
To minimize measurement errors, shield inputs
from extern al noise when measuring low-voltage,
low-frequency signals
Input Characteristics
Uncertainty
90 d ays 1 year
Range Frequency
23 °C ± 5 °C 23 ° C ±5 °C
Temperature C oef ficien t/°C
Outside 18 - 28 °C
20 Hz - 2 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
2 kHz - 20 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
20 kHz - 200 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
100mVto750V*
4, 5
200 kHz - 1 MHz 0.01 + 0.004 0.01 + 0.006 0.002 + 0.002
*
4
Input >100 mV.
*
5
Limitedto8×10
7
VHz.
Continuity
Characteristic
Description
Continuity Threshold 20 Ω
Test Current
1mA
Response Time
100 S/s with audible tone
Rate Fast
Maximum Reading
199.99 Ω
Resolution
0.01 Ω
Diode Test
Characteristic
Description
Response Time
100 S/s with audible tone
Rate Fast
Maximum Reading 1.9999 V
Resolution 0.1 mV
10 www.tektronix.com/dmm