User's Manual
Functional Check
18
P6330 3.5 GHz Differential Probe Instruction Manua l
NOTE. If your instrument supports probe calibration routines, now is
a good time to perform them.
6. Disconnect the probe from the PROB E COMPENSATION
connector and connect the two KlipChips t ogether.
7. With the probe offset set to 0.0 V, the oscilloscope display should
be at the ground reference.
8. Set the oscilloscope volts/division to 500 mV.
9. Adjust the probe offset. The displayed wave form should vary
between approximately +1.0 V and --1.0 V.