Stereo System User Manual

Displaying Waveforms
3-62
CSA8000B & TDS8000B User Manual
Use the procedure that follows to become familiar with the display adjustments
you can make.
Overview To control the Main view Related control elements and resources
Prerequisites 1. The instrument must be installed with sampling modules
in place.
2. The acquisiti on system should be set to run
continuously.
3. Also, an appropriate trigger signal must be applied to
the instrum ent and triggering must be set up.
See the sampling Module user manuals for sampling
module installation. See page 3--24 for acquisition
setup and page 3--48 for trigger setup in this manual.
Set the vertical
display
parameters
4. Push a Vertical Source butt on (turns amber) to
assign the numbered buttons 1--8 t o operate on
channel, reference, or m ath wavef orms. Push a
numbered button 1--8 to select the waveform (it
displays).
A waveform button lights when its waveform is on:
H Lighted green: waveform is on but not
selected
H Lighted amber: waveform is on and selected
Hint. Step 4 assumes any reference or math
waveforms you select are defi ned. See Table 3--2
on page 3--56 if you need help defining t hese
waveforms.
5. Use the Vert ical knobs t o achieve a good display
of each waveform you select .
To Display Waveforms in
the Main Time Base View