User manual

1760 Series Component Waveform/Vector Monitor User Manual 2-11
Operator’s Checkout
Procedure
Before proceeding, read At A Glance startingonpage2-1.
This procedure is designed for operator familiarization and
for checking basic instrument operation (not measurement
quantities or specifications). All illustrations are of the 1765
unless otherwise note d.
If performing this procedure re veals improper instrument
operation, first che ck the operation of the associated equip-
ment. If the associated equipment is operating normally,
refer the 1760--Series to qualified service personnel for
repair or adjustment.
Required Equipment
The following equipment is re quired to perform this proce-
dure:
1. Television Signal Generator with:
Composite Color Bars
Component signal (Y, B--Y, R--Y)
Channel Timing (Bowtie)
Black Burst or comparable signal with burst and sync.
For example: TEKTRONIX TSG 130--Series Multifor-
mat Signal Gene rator
2. Coaxial Cable, 75 (5)
For example: 42-inch RG59U (Tektronix Part No.
012-0159-00)
3. 75 Terminators, End-line (4)
For example: (Tektronix Part No. 011-0102-00)