User manual

Installation
1760 Series Component Waveform/Vector Monitor User Manual
1-10
2. Package the instrument in the original packaging materi-
als. If the original packaging materials are not available,
follow these direc tions:
a. Obtain a cart on of corrugated cardboard having
inside dimensions six or more inches greater than
the dimensions of the instrument. Use a shipping
carton that has a test strength of at least 275 pounds.
b. Surround the instrume nt wit h a protective bag
(anti-static preferred). If the instrument is not in a
cabinet, wrap a cardboard pie ce around the bagged
instrument to protect components.
c. Pack dunnage or urethane foam be tween the instru-
ment and the carton. If using Styrofoam kernels,
overfill the box and compress by closing the lid.
There should be three inches of tightly packed
cushioning on al l sides of the instrument.
3. Seal the carton with shipping tape, industrial stapler, or
both.
Electrical Installation
P ower Source
These monitors are designed to operate from a single-phase
power source having one of its current-carrying conductors at
or near earth ground (the neutral conductor). Only the line
conductor is fused for over-current protection. Systems that
have both current-carrying conductors l ive with respect to
ground (such as phase-to-phase on multiphase systems) are
not recommended a s power sources. A protective ground
connection by way of the grounding conductor in the power
cord is essential for safe operation.