User's Manual
Table Of Contents
- Title page
- Table of Contents
- General Safety Summary
- Preface
- Getting Started
- Operating Basics
- Reference
- Reference
- Menu Structures
- The Setup Menu Screen
- The Graphical Waveform Editor
- The Pattern Editor
- Quick Editing
- The Table Editor
- The Equation Editor
- The Sequence Editor
- The APPL Menu
- The UTILITY Window
- External Keyboards
- Setting General Purpose Knob Direction
- Formatting a Floppy Disk
- Displaying Disk Usage
- Screen Display Enable/Disable
- Focused Color
- Displaying Instrument Status
- Internal Clock (Date and Time)
- Resetting the Instrument
- Connecting to a GPIB Network
- Ethernet Networking
- Hardcopy
- Calibration and Diagnostics
- Upgrading the System Software
- Capturing Waveforms
- Waveform Programming Language
- Command Descriptions
- Programming Examples
- File Conversion
- File Management
- FG Mode
- Waveform Mixing Mode
- Synchronous Operation Mode (AWG710B only)
- Appendices
- Appendix A: Specifications (AWG710B)
- Appendix A: Specifications (AWG710)
- Appendix B: Performance Verification (AWG710B)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- External Clock Input and VCO Out Output Tests
- VCO OUT Output Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Synchronous Operation Tests
- Appendix B: Performance Verification (AWG710)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- 1/4 Clock Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Appendix C: Inspection and Cleaning
- Appendix D: Sample Waveforms
- Appendix E: File Transfer Interface Outline
- Appendix F: Miscellaneous
- Appendix G: Sequence File Text Format
- Index

Appendix B: Performance Verification (AWG710)
B-98 AWG710&AWG710B Arbitrary Waveform Generator User Manual
1/4 Clock Frequency and 10 MHz Reference Input Tests
These procedures check the 10 MHz reference input function of the AWG710
Arbitrary Waveform Generator.
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50 Ω SMA coaxial cable and SMA(Fe)–BNC(Ma) adapter to connect the
AWG710 Arbitrary Waveform Generator 1/4 CLOCK OUT connector to the
input A connector on the frequency counter.
2. Use a 50 Ω BNC coaxial cable to connect the AWG710 Arbitrary Waveform
Generator 10 MHz REF IN connector to the function generator output
connector (see Figure B-62).
Figure B-62: 1/4 Clock frequency and 10 MHz reference input initial test hookup
a. Set the frequency counter controls as follows:
Equipment
required
A 50 Ω SMA coaxial cable, A 50 Ω BNC coaxial cable, a
SMA(Fe)–BNC(Ma) adapter, a frequency counter, and a function
generator.
Prerequisites
The AWG710 Arbitrary Waveform Generator must meet the prerequisites
listed on page B-61.
Function Generator (AFG310)
AWG710 Arbitrary Waveform
Generator rear panel
Frequency Counter
50 Ω BNC coaxial cable 50 Ω SMA coaxial cable
10 MHz REF IN
1/4 CLOCK OUT
SMA(Female)-BNC
(Male) adapter
INPUT A
Coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . AC
FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A FREQ
Gate time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.2 s
Trigger Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 V