User's Manual
Table Of Contents
- Title page
- Table of Contents
- General Safety Summary
- Preface
- Getting Started
- Operating Basics
- Reference
- Reference
- Menu Structures
- The Setup Menu Screen
- The Graphical Waveform Editor
- The Pattern Editor
- Quick Editing
- The Table Editor
- The Equation Editor
- The Sequence Editor
- The APPL Menu
- The UTILITY Window
- External Keyboards
- Setting General Purpose Knob Direction
- Formatting a Floppy Disk
- Displaying Disk Usage
- Screen Display Enable/Disable
- Focused Color
- Displaying Instrument Status
- Internal Clock (Date and Time)
- Resetting the Instrument
- Connecting to a GPIB Network
- Ethernet Networking
- Hardcopy
- Calibration and Diagnostics
- Upgrading the System Software
- Capturing Waveforms
- Waveform Programming Language
- Command Descriptions
- Programming Examples
- File Conversion
- File Management
- FG Mode
- Waveform Mixing Mode
- Synchronous Operation Mode (AWG710B only)
- Appendices
- Appendix A: Specifications (AWG710B)
- Appendix A: Specifications (AWG710)
- Appendix B: Performance Verification (AWG710B)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- External Clock Input and VCO Out Output Tests
- VCO OUT Output Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Synchronous Operation Tests
- Appendix B: Performance Verification (AWG710)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- 1/4 Clock Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Appendix C: Inspection and Cleaning
- Appendix D: Sample Waveforms
- Appendix E: File Transfer Interface Outline
- Appendix F: Miscellaneous
- Appendix G: Sequence File Text Format
- Index

Appendix B: Performance Verification (AWG710)
B-74 AWG710&AWG710B Arbitrary Waveform Generator User Manual
5. Do the following substeps to set the AWG710 Arbitrary Waveform Generator
amplitude and confirm the offset setting:
a. Push VERTICAL MENU (front–panel)!Amplitude (side).
b. Push 0, ., 0, 2 and ENTER keys in this order or turn the general purpose
knob to set the amplitude to 0.020 V.
c. Verify that the offset setting display on the Offset side button is 0.000 V.
If the offset display is not set correctly, push the Offset side button, and push
0 and then ENTER key.
6. Do the following substeps to check the amplitude accuracy of a 20 mV
amplitude setting:
a. Write the DMM reading as a positive voltage.
b. Push the FORCE EVENT button.
c. Write the DMM reading as a negative voltage.
d. Verify that the positive minus negative voltages fall within
20 mV ± 2.4 mV.
e. Push 0, ., 2 and ENTER keys in this order or turn the general purpose knob
to set the amplitude to 0.200 V.
7. Do the following to check the amplitude accuracy of 200 mV amplitude
setting:
a. Push the FORCE EVENT button.
b. Write the DMM reading as a positive voltage.
c. Push the FORCE EVENT button.
d. Write the DMM reading as a negative voltage.
e. Verify that the positive minus negative voltages fall within
200 mV ± 6 mV.
f. Push the 2 and the ENTER keys in this order or turn the general purpose
knob to set the amplitude to 2 V.
8. Do the following substeps to check the amplitude accuracy of 2 V amplitude
setting:
a. Push the FORCE EVENT button.
b. Write the DMM reading as a positive voltage.
c. Push the FORCE EVENT button.