User's Manual
Table Of Contents
- Title page
- Table of Contents
- General Safety Summary
- Preface
- Getting Started
- Operating Basics
- Reference
- Reference
- Menu Structures
- The Setup Menu Screen
- The Graphical Waveform Editor
- The Pattern Editor
- Quick Editing
- The Table Editor
- The Equation Editor
- The Sequence Editor
- The APPL Menu
- The UTILITY Window
- External Keyboards
- Setting General Purpose Knob Direction
- Formatting a Floppy Disk
- Displaying Disk Usage
- Screen Display Enable/Disable
- Focused Color
- Displaying Instrument Status
- Internal Clock (Date and Time)
- Resetting the Instrument
- Connecting to a GPIB Network
- Ethernet Networking
- Hardcopy
- Calibration and Diagnostics
- Upgrading the System Software
- Capturing Waveforms
- Waveform Programming Language
- Command Descriptions
- Programming Examples
- File Conversion
- File Management
- FG Mode
- Waveform Mixing Mode
- Synchronous Operation Mode (AWG710B only)
- Appendices
- Appendix A: Specifications (AWG710B)
- Appendix A: Specifications (AWG710)
- Appendix B: Performance Verification (AWG710B)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- External Clock Input and VCO Out Output Tests
- VCO OUT Output Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Synchronous Operation Tests
- Appendix B: Performance Verification (AWG710)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- 1/4 Clock Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Appendix C: Inspection and Cleaning
- Appendix D: Sample Waveforms
- Appendix E: File Transfer Interface Outline
- Appendix F: Miscellaneous
- Appendix G: Sequence File Text Format
- Index

Appendix B: Performance Verification (AWG710)
B-60 AWG710&AWG710B Arbitrary Waveform Generator User Manual
Performance Tests
This section contains a collection of procedures for checking that the AWG710
Arbitrary Waveform Generator performs as warranted.
The procedures are arranged in eight logical groupings, presented in the following
order:
The performance tests check all of the characteristics that are designated as
checked in Appendix A:Specifications. (The characteristic items that must be
checked are listed with the check mark (n) in Appendix A: Specifications).
Table B-4: Performance test items
Titles
See
(performance
verification) Test items See (specifications)
Operating mode tests Page B-66 Continuous, Triggered, and Gated
mode normality
Page A-28
Amplitude and offset accuracy tests (normal out) Page B-72 Amplitude accuracy, and DC offset
accuracy
Page A-29
Amplitude, DC offset and rise time accuracy tests
(direct DA out)
Page B-77 Amplitude accuracy, DC offset
accuracy and Rise time accuracy
Page A-30
Amplitude, DC offset and
Rise time accuracy test (Option2)
Page B-81 Amplitude accuracy, DC offset
accuracy and Rise time accuracy
Page A-30
Pulse response tests (normal out) Page B-85 Rise time accuracy Page A-30
Trigger input tests Page B-87 Trigger level accuracy, and Trigger
function normality
Page A-33
Event input and enhanced mode tests Page B-91 External event input function normality,
and
Event mode normality
Page A-35
1/4 Clock frequency and 10 MHz reference input
tests
Page B-98 1/4 clock output frequency and
Reference input normality accuracy
Page A-31
Marker output tests Page B-100 Marker output level accuracy Page A-31
NOTE. These procedures extend the confidence level provided by the basic
procedures described on page B-56. The basic procedures should be done first, and
then these procedures performed if desired.