User's Manual
Table Of Contents
- Title page
- Table of Contents
- General Safety Summary
- Preface
- Getting Started
- Operating Basics
- Reference
- Reference
- Menu Structures
- The Setup Menu Screen
- The Graphical Waveform Editor
- The Pattern Editor
- Quick Editing
- The Table Editor
- The Equation Editor
- The Sequence Editor
- The APPL Menu
- The UTILITY Window
- External Keyboards
- Setting General Purpose Knob Direction
- Formatting a Floppy Disk
- Displaying Disk Usage
- Screen Display Enable/Disable
- Focused Color
- Displaying Instrument Status
- Internal Clock (Date and Time)
- Resetting the Instrument
- Connecting to a GPIB Network
- Ethernet Networking
- Hardcopy
- Calibration and Diagnostics
- Upgrading the System Software
- Capturing Waveforms
- Waveform Programming Language
- Command Descriptions
- Programming Examples
- File Conversion
- File Management
- FG Mode
- Waveform Mixing Mode
- Synchronous Operation Mode (AWG710B only)
- Appendices
- Appendix A: Specifications (AWG710B)
- Appendix A: Specifications (AWG710)
- Appendix B: Performance Verification (AWG710B)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- External Clock Input and VCO Out Output Tests
- VCO OUT Output Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Synchronous Operation Tests
- Appendix B: Performance Verification (AWG710)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- 1/4 Clock Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Appendix C: Inspection and Cleaning
- Appendix D: Sample Waveforms
- Appendix E: File Transfer Interface Outline
- Appendix F: Miscellaneous
- Appendix G: Sequence File Text Format
- Index

Appendix B: Performance Verification (AWG710B)
B-46 AWG710&AWG710B Arbitrary Waveform Generator User Manual
External Clock Input and VCO Out Output Tests
These procedures check the external clock input function and the VCO output
function of the AWG710B Arbitrary Waveform Generator.
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50 Ω SMA coaxial cable and an SMA(Fe)–BNC(Fe) adapter to connect
the AWG710B Arbitrary Waveform Generator VCO OUT output connector to
the oscilloscope CH1 input connector.
2. Use a 50 Ω SMA coaxial cable to connect the AWG710B Arbitrary Waveform
Generator EXT CLOCK IN input connector to the signal generator output
connector.
Figure B-31: Trigger input initial test hookup
3. Set the oscilloscope controls as follows:
Equipment
required
Two 50 Ω SMA coaxial cables, a SMA(Fe)–BNC(Ma) adapter, a signal
generator, and an oscilloscope (TDS700).
Prerequisites
The AWG710B Arbitrary Waveform Generator must meet the
prerequisites listed on page B-7.
Signal Generator
AWG710B Arbitrary Waveform
Generator rear panel
Oscilloscope (TDS700)
50 Ω SMA coaxial cable
VCO OUTOutput
CH1 input
SMA(Female)-BNC
(Male) adapter
SMA(Female)-BNC
(Male) adapter
50 Ω SMA coaxial cable
EXT CLOCK IN
Vertical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CH1
CH1 coupling. . . . . . . . . . . . . . . . . . . . . . . . . DC
CH1 scale . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 mV/div
CH1 input impedance . . . . . . . . . . . . . . . . . . 50 Ω
Horizontal