User's Manual
Table Of Contents
- Title page
- Table of Contents
- General Safety Summary
- Preface
- Getting Started
- Operating Basics
- Reference
- Reference
- Menu Structures
- The Setup Menu Screen
- The Graphical Waveform Editor
- The Pattern Editor
- Quick Editing
- The Table Editor
- The Equation Editor
- The Sequence Editor
- The APPL Menu
- The UTILITY Window
- External Keyboards
- Setting General Purpose Knob Direction
- Formatting a Floppy Disk
- Displaying Disk Usage
- Screen Display Enable/Disable
- Focused Color
- Displaying Instrument Status
- Internal Clock (Date and Time)
- Resetting the Instrument
- Connecting to a GPIB Network
- Ethernet Networking
- Hardcopy
- Calibration and Diagnostics
- Upgrading the System Software
- Capturing Waveforms
- Waveform Programming Language
- Command Descriptions
- Programming Examples
- File Conversion
- File Management
- FG Mode
- Waveform Mixing Mode
- Synchronous Operation Mode (AWG710B only)
- Appendices
- Appendix A: Specifications (AWG710B)
- Appendix A: Specifications (AWG710)
- Appendix B: Performance Verification (AWG710B)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- External Clock Input and VCO Out Output Tests
- VCO OUT Output Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Synchronous Operation Tests
- Appendix B: Performance Verification (AWG710)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- 1/4 Clock Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Appendix C: Inspection and Cleaning
- Appendix D: Sample Waveforms
- Appendix E: File Transfer Interface Outline
- Appendix F: Miscellaneous
- Appendix G: Sequence File Text Format
- Index

Appendix B: Performance Verification (AWG710B)
B-22 AWG710&AWG710B Arbitrary Waveform Generator User Manual
d. Push 0, ., 0, 2 and ENTER keys in this order or turn the general purpose
knob to set the amplitude to 0.020 V.
3. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
4. Do the following substeps to set the AWG710B Arbitrary Waveform
Generator offset:
a. Push the Offset side button.
b. Push 0 and ENTER keys in this order.
c. Verify that the reading on the DMM falls within 0 V ± 10 mV.
d. Push 0, ., 5 and ENTER keys in this order to change the AWG710B
Arbitrary Waveform Generator offset to 0.5 V.
e. Verify that the reading on the DMM falls within 0.5 V ± 0.0175 V.
f. Push –, 0, ., 5 and ENTER keys in this order to change the AWG710B
Arbitrary Waveform Generator offset to –0.5 V.
g. Check that the reading on the DMM falls within –0.5 V ± 0.0175 V.
5. Do the following substeps to change the connection to check the CH1
:
a. Push the CH1 output button.
b. Disconnect the adapter, terminator and cable from the CH1 output
connector.
c. Connect the adapter, terminator, and cable (removed in step b) to the CH1
output connector.
d. Push the CH1
output button to turn on the CH1 LED.
6. Repeat the Check Offset Accuracy procedure for the AWG710B Arbitrary
Waveform Generator CH1
.
7. Push the CH1
output button to turn off the CH1 LED.
8. Disconnect the test hookup from the CH1
output connector.