User's Manual
Table Of Contents
- Title page
- Table of Contents
- General Safety Summary
- Preface
- Getting Started
- Operating Basics
- Reference
- Reference
- Menu Structures
- The Setup Menu Screen
- The Graphical Waveform Editor
- The Pattern Editor
- Quick Editing
- The Table Editor
- The Equation Editor
- The Sequence Editor
- The APPL Menu
- The UTILITY Window
- External Keyboards
- Setting General Purpose Knob Direction
- Formatting a Floppy Disk
- Displaying Disk Usage
- Screen Display Enable/Disable
- Focused Color
- Displaying Instrument Status
- Internal Clock (Date and Time)
- Resetting the Instrument
- Connecting to a GPIB Network
- Ethernet Networking
- Hardcopy
- Calibration and Diagnostics
- Upgrading the System Software
- Capturing Waveforms
- Waveform Programming Language
- Command Descriptions
- Programming Examples
- File Conversion
- File Management
- FG Mode
- Waveform Mixing Mode
- Synchronous Operation Mode (AWG710B only)
- Appendices
- Appendix A: Specifications (AWG710B)
- Appendix A: Specifications (AWG710)
- Appendix B: Performance Verification (AWG710B)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- External Clock Input and VCO Out Output Tests
- VCO OUT Output Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Synchronous Operation Tests
- Appendix B: Performance Verification (AWG710)
- Conventions
- Self Tests
- Performance Tests
- Operating Mode Tests
- Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (Direct DA Out), (except option 02)
- Amplitude, Offset Accuracy and Rise Time Tests (for option 02)
- Pulse Response Tests (Normal Out), (except option 02)
- Trigger Input Tests
- Event Input and Enhanced Mode Tests
- 1/4 Clock Frequency and 10 MHz Reference Input Tests
- Marker Output Tests
- Appendix C: Inspection and Cleaning
- Appendix D: Sample Waveforms
- Appendix E: File Transfer Interface Outline
- Appendix F: Miscellaneous
- Appendix G: Sequence File Text Format
- Index

The Graphical Waveform Editor
3-60 AWG710&AWG710B Arbitrary Waveform Generator User Manual
The instrument replaces, inserts, adds, or multiplies the edit area with the
specified standard waveform data.
Figure 3-9: Standard Function Waveform dialog box
Table 3-10: Standard Function Waveform dialog box parameters
Parameter Description
Type Specifies the type of standard function waveform to create. You can select Sine,
Triangle, Square, Ramp, DC, Gaussian Noise, or Random Noise.
Operation Selects how the standard function waveform is added to the edit area.
Replace replaces the edit area data with the specified standard function
waveform. This operation does not change the waveform data record length.
Insert inserts the standard function waveform at the active cursor position. This
operation increases the waveform data record length by the amount of the
inserted waveform.
Add replaces the edit area data with the sum of the current edit area data and
the specified standard function waveform. This operation does not change the
waveform data record length.
Mul replaces the edit area data with the product of the current edit area data and
the standard function waveform. This operation does not change the waveform
data record length.
Cycle Specifies the number of function waveform cycles to insert in the specified cursor
area. The range of values is from 0.1 to 100,000 in 0.000001 to 0.001
increments. The default value is 1 cycle.
If the Operation field is set to Replace, Add, or Mul, the Cycle field value
determines the Frequency field value according to the equation
Frequency = Cycle x clock frequency / data length.
Frequency Specifies the frequency of the function waveform to insert in the specified cursor
area. The range of values is from 0.1 Hz to 500 MHz, with 9–digit accuracy.
If the Operation field is set to Replace, Add, or Mul, the Frequency field
determines the Cycle field value according to the equation
Cycle = Frequency x data length / clock frequency.