Datasheet

Sweep
Type Linear, Logarithmic
Waveforms All, except Pulse, Noise, DC and PRBS
Sweep time 1 ms to 500 s
Hold/return time 0 s to 500 s
Maximum total sweep time 500 s
Accuracy, typical: ≤ 0.4%
Minimum start/stop frequency All except ARB: 1 μHz
ARB: 1 mHz
Maximum start/stop frequency
AFG31021 /
AFG31022
AFG31051 /
AFG31052
AFG31101 /
AFG31102
AFG31151 /
AFG31152
AFG31251 /
AFG31252
Sine 25 MHz 50 MHz 75 MHz 125 MHz 250 MHz
Square, Arb 20 MHz 40 MHz 80 MHz 120 MHz 160 MHz
Others 500 kHz 800 kHz 1 MHz 1 MHz 2.5 MHz
Burst
Waveform All except Noise, DC and PRBS
Type Triggered, gated
Burst count 1 to 1,000,000 cycles or Infinite
Intenal trigger rate 1 μs to 500.0 s
Gate and trigger sources Internal, external, remote interface
InstaView
Waveforms All except noise and PRBS
Run mode Continuous in Basic mode
Maximum measurement range
(DC + peak AC voltage)
-10 V to 10 V
DC level measurement
Specification
Characteristic
Accuracy (into 50Ω), typical ± (2 % of |setting| + 20 mVpp)
Resolution 1 mV or 4 digits
Amplitude measurement
Specification
Characteristic
Accuracy (sine, 1 Vp-p, into
50 Ω), typical)
± (2 % of setting + 20 mV)
Resolution 1 mV or 4 digits
Bandwidth (-3 dB)
AFG31021 / AFG31022
AFG31051 / AFG31052 AFG31101 / AFG31102 AFG31151 / AFG31152 AFG31251 / AFG31252
500 MHz TBD TBD
AFG31000 Series
General characteristics - Basic mode
www.tektronix.com 11