Datasheet

System characteristics
Output Frequency Resolution
Frequency accuracy ±10
-6
of setting (all except ARB), 0 °C to 50 °C (32 °F to 122 °F)
±10
-6
of setting ± 1 μHz (ARB), 0 °C to 50 °C (32 °F to 122 °F)
Aging ±1.0 x 10
-6
per year
Phase
Range -180° to +180°
Resolution 0.01° (sine)
0.1° (other waveforms)
Internal noise add
Level 0.0% to 50% of amplitude (V
P-P
) setting
Resolution 1%
Remote program interface GPIB, Ethernet 10BASE-T / 100BASE-TX / 1000BASE-T, USB 2.0
Maximum configuration times
USB
LAN GPIB
Function change 61 ms 61 ms 63 ms
Frequency change (except Pulse) 3 ms 4 ms 6 ms
Frequency change (Pulse) 2.5 ms 3 ms 8 ms
Amplitude change 65 ms 66 ms 77 ms
Select user ARB (4k points from USB Memory) 43 ms 40 ms 53 ms
Select user ARB (128k points from USB Memory) 86 ms 92 ms 92 ms
Data download time for 4k points 36 ms 21 ms 21 ms
Power source
Source 100-240 V, 47-63 Hz
115 V, 360-440 Hz
Consumption 120 W
Warm up time, typical 20 minutes minimum
Power on self diagnosis time < 24 s
Acoustic noise < 50 dBA
Display 9-inch capacitive touch screen with 800 * 480 resolution
User interface and Help languages English, French, German, Japanese, Korean, Portuguese, Simplified and Traditional Chinese, Russian (user selectable)
Datasheet
14 www.tektronix.com