Datasheet

Sweeping
Modulation, sweeping, and burst modes are only available for channel 1 on the AFG1022.
The AFG1062 supports equal strong channels with modulation, sweeping, and burst modes.
Carrier waveforms Sine, square, ramp, arbitrary (AFG1062 only)
Minimum start-stop frequency 1 μHz
Maximum start-stop frequency
Sine
AFG1022 AFG1062
25 MHz 60 MHz
Square
12.5 MHz
30 MHz
Ramp
1 MHz
2 MHz
Type Linear, logarithmic
Direction Up / down
Sweep time 1 ms to 500 s ± 0.1%
Trigger sources Internal, external, or manual
Burst
Modulation, sweeping, and burst modes are only available for channel 1 on the AFG1022.
The AFG1062 supports equal strong channels with modulation, sweeping, and burst modes.
Waveforms Sine, square, ramp, pulse, arbitrary except DC and noise
Types AFG1022: count (1 to 50,000 cycles), infinite, gated
AFG1062: count (1 to 1,000,000 cycles), infinite, gated
Start phase -360° to +360°
Trigger sources Internal, external, or manual
Internal trigger interval (40 ns or (cycles x period) to 500 s) ± 1%
Gate source External trigger
Frequency counter
Function Frequency, period, positive pulse width, duty cycle
Frequency range 100 mHz to 200 MHz
Frequency resolution 6 digits
Coupling mode AC, DC
Datasheet
6 www.tektronix.com