Datasheet
RF Power level trigger
Frequency range Models with option SA3: 1 MHz to 3 GHz
Models with option SA6: 1 MHz to 3.75 GHz; 2.75 GHz to 4.5 GHz, 3.5 GHz to 6.0 GHz
Amplitude operating level 0 dB to -30 dB from Reference Level
Amplitude range +10 dB to -40 dB from Reference Level and within the range of -65 dBm to +30 dBm
Minumum pulse generation 10 µs On Time with a minimum settling Off Time of 10 µs
Spectrum analyzer to analog
channel skew
<5 ns
RF acquisition length
Span
Maximum RF acquisition time
>2 GHz 5 ms
>1 GHz - 2 GHz 10 ms
>800 MHz - 1 GHz 20 ms
>500 MHz - 800 MHz 25 ms
>400 MHz - 500 MHz 40 ms
>250 MHz - 400 MHz 50 ms
>200 MHz - 250 MHz 80 ms
>160 MHz - 200 MHz 100 ms
>125 MHz - 160 MHz 125 ms
<125 MHz 158 ms
FFT window types, factors, and
RBW accuracy
FFT window
Factor RBW accuracy
Kaiser 2.23 0.90%
Rectangular 0.89 2.25%
Hamming 1.30 1.54%
Hanning 1.44 1.39%
Blackman-Harris 1.90 1.05%
Flat-Top 3.77 0.53%
3- Arbitrary Function Generator (requires Option MDO4AFG)
Waveforms Sine, Square, Pulse, Ramp/Triangle, DC, Noise, Sin(x)/x (Sinc), Gaussian, Lorentz, Exponential Rise, Exponential Decay,
Haversine, Cardiac, and Arbitrary.
Sine
Frequency range 0.1 Hz to 50 MHz
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
Amplitude flatness (typical) ±0.5 dB at 1 kHz (±1.5 dB for <20 mV
p-p
amplitudes)
Total harmonic distortion
(typical)
1% into 50 Ω
2% for amplitude < 50 mV and frequencies > 10 MHz
3% for amplitude < 20 mV and frequencies > 10 MHz
Spurious free dynamic range
(SFDR) (typical)
-40 dBc (V
p-p
≥ 0.1 V); -30dBc (V
p-p
≤ 0.1 V), 50 Ω load
Square / Pulse
Frequency range 0.1 Hz to 25 MHz
Amplitude range 20 mV
p-p
to 5 V
p-p
into Hi-Z; 10 mV
p-p
to 2.5 V
p-p
into 50 Ω
MDO4000C Series Oscilloscope
2- Spectrum Analyzer (requires Option SA3 or SA6)
www.tek.com 29