Datasheet

Absolute amplitude accuracy
(AAA) and channel response (CR),
with TPA-N-PRE preamp attached
AAA: ≤ ±0.5 dB (typical), 18 C - 28 C temperature range, 50 kHz to 6 GHz, either preamp state.
AAA: ≤ ±2.0 dB (typical), 18 C - 28 C temperature range, 9 kHz to 50 kHz, either preamp state.
AAA: ≤ ±2.3 dB (typical), over full operating range, either preamp state.
CR: 0.0 dB
Crosstalk to spectrum analyzer
from oscilloscope channels
≤1 GHz input frequencies < -68 dB from ref level
>1 GHz - 2 GHz input
frequencies
< -48 dB from ref level
Phase noise at 1 GHz CW
1 kHz < -104 dBc/Hz,(typical)
10 kHz < -108 dBc/Hz, < -111 dBc/Hz (typical)
100 kHz < -110 dBc/Hz, < -113 dBc/Hz (typical)
1 MHz < -120 dBc/Hz, < -123 dBc/Hz (typical)
Reference frequency error
(cumulative)
Cumulative error: 1.6 x 10
-6
Includes allowances for aging per year, reference frequency calibration accuracy, and temperature stability
Valid over the recommended one-year calibration interval, from 0 °C to +50 °C
Marker frequency measurement
accuracy
±((1.6x 10
-6
x Marker-Frequency) + (0.001 x span + 2)) Hz
Example: assuming the span is set to 10 kHz and the marker is at 1500 MHz, this would result in a Frequency Measurement
Accuracy of +/-((1.6 x 10 -6 x 1500 MHz) + (0.001 x 10 kHz + 2)) = +/- 2.412 kHz.
Marker Frequency with Span/RBW ≤ 1000:1
Reference Frequency Error with Marker level to displayed noise level > 30 dB
Frequency measurement
resoultion
1 Hz
Maximum operating input level
Average continuous power +30 dBm (1 W) for reference levels ≥ -20 dBm
+24 dBm (0.25 W) for reference levels < -20 dBm
DC maximum before damage ±40 V
DC
Maximum power before
damage (CW)
+32 dBm (1.6 W) for reference levels ≥ -20 dBm
+25 dBm (0.32 W) for reference levels of < -20 dBm
Maximum power before
damage (pulse)
Peak Pulse Power: +45 dBm (32 W)
Peak Pulse Power defined as <10 μs pulse width, <1% duty cycle, and reference level of ≥ +10 dBm
Maximum operating input level
with TPA-N-PRE preamp attached
Average continuous power +30 dBm (1 W)
DC maximum before damage ±20 V
DC
Maximum power before
damage (CW)
+30 dBm (1 W)
Maximum power before
damage (pulse)
+45 dBm (32 W) (<10 μs pulse width, <1% duty cycle, and reference level of ≥ +10 dBm)
Datasheet
2- Spectrum Analyzer (requires Option SA3 or SA6)
28 www.tek.com