User manual

A622 Instructions
3
Operating Basics
Before using the probe, the batte ries must be installed. See the
battery installation instructions on page 7.
WARNING. Do not clamp the probe onto circuits with voltages greater
than 600 VAC. Personal injury or damage to the probe may result.
Always connect the probe to the instrument before clamping onto the
circuit under test.
1. Connect the probe BNC connector to the oscilloscope input. Start
by setting the oscilloscope channel vertical coupling to DC volts,
the vertical deflection to 0.1 V/div.
2. Move the OFF/Range switchtothe10 mV/A or 100 mV/A
position to power on the probe.
The A622 current probe has a green LED power/battery indicator.
If the LED does not light, replace the battery.
3. Use the ZERO adjustme nt t o ze ro or offset the probe output.
4. Connect the probe to the circuit by opening the ja ws and
clamping around the conductor. See Figure 2.
NOTE. Clamping around both the “hot” and neutral wires may give
you a zero reading.
(Remember to unclamp the probe from the conductor before
disconnecting it from your meter or instrument.)