User manual
Performance Veri fication
A621 Instructions
15
7. Check that the waveform peak to peak amplitude measures
1.4 ±0.1 V
P--P
on the oscilloscope.
8. Continue to che ck the probe accuracy using the parameters in
Table 6.
Table 6: AC Accuracy Check Parameters
Step
Probe Range Oscilloscope
V/div setting
Multimeter
reading
Oscilloscope dis-
play
1. 100 mV/A 200 mV/div . 50 ±1mA 1.4 ±0.1 V
P-- P
2. 10 mV/A 20 mV/div. 50 ±1mA 140 ±10 mV
P-- P
3. 1mV/A 2mV/div. 50 ±1mA 14 ±1mV
P-- P