Performance Veri fication
14
A621 Instructions
AC Accuracy Check
1. Connect the 100 turn current loop, the mu ltimeter current inputs,
and the signal genera tor output in series, using the patch cords
and adapter. Set the multimeter to display AC
RMS
current. See
Figure 5.
Oscilloscope
Multimeter
Signal
Generator
A621
100 Turn coil
Figure 5: Setup for Maximum Input Current and Accuracy Checks
2. Adjust the signal generator output to between 49 mA
RMS
and
51 mA
RMS
as measured on the multimeter.
3. Connect the A621 to the oscilloscope, and make the following
oscilloscope settings:
H time base to 5 mS / div.
H vertical deflection to 500 mV / div.
H couplingtoDC
4. Set the A621 sensitivity switch to 100 mV / A and center the
trace on the oscilloscope screen.
5. Clamp the A621 around the 100 turn current loop.
6. If necessary, readjust the signal gene rator output to read betwee n
49 mA
RMS
and51mA
RMS
on the multimeter.