Performance Veri fication
A621 Instructions
13
Table 5: Test Equipment Required for Performance Verification
Item
Minimum Requirements Recommended Example
Oscilloscope
1MΩ input, general purpose
oscilloscope
Tektronix
TAS/TDS series
Signal Generator 60 Hz Sine Wave, ≥50 mA
RMS
into <1 Ω
Tektronix FG 501A
1
,
FG 504
1
or equivalent
Multimeter ≥0.5 A max. input current,
accuracy ≤0.6% of reading on
AC Current measurement at
60 Hz
Tektronix DM 504A
1
or equivalent
Current Loop, 100 turn 100 turns coated wire,
20 AWG (≈0.9 mm),
loop diameter 7.5 cm (3 in),
alligator clips on ends
See procedure
18 inch banana jack-to-banana jack patch cords, three
required
Tektronix part numbers:
012-0031-00
012-0039-00
1
Requires Tektronix TM 500 Series or Tektronix TM 5000 Series Power Module
100 Turn Current Loop
CAUTION. It is not recommended that a coil with greater than 100
turns be made. Magnetic fields are produced that may cause a
malfunction in heart pacemakers, or damage to sensitive equipment.
Coil currents in excess of 0.2 Amperes are not recommended.
For the accuracy check you will need a simple current loop. Wind
100 turns of 20 AWG (≈0.9m diameter) coated wire around a
cylindrical form approximately 7.5 cm (3 inches) in diameter.
NOTE. Be sure the current loop has exactly 100 turns. A significant
error will result for each extra or missing turn.