Datasheet

LOW LEVEL MEASURE & SOURCE
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY
(U.S. only)
3390
50MHz Arbitrary Waveform/
Function Generator
GENERAL
POWER SUPPLY: CAT II 110–240VAC ±10%.
POWER CORD FREQUENCY: 50Hz to 60Hz.
POWER CONSUMPTION: 50VA max.
OPERATING ENVIRONMENT: 0° to 50°C.
STORAGE TEMPERATURE: –30° to 70°C.
INTERFACE: USB, LAN, LXI-C, GPIB.
LANGUAGE: SCPI-1993, IEEE-488.2.
DIMENSIONS: 107mm high × 224mm wide × 380mm
deep (4.2 in. × 8.8 in. × 15 in.).
WEIGHT: 4.08kg.
SAFETY: Conforms with European Union Directive
73/23/EEC, EN 61010-1.
EMC: Conforms with European Union Directive
89/336/EEC, EN 61326-1.
WARM-UP: 1 hour.
NOTES
1. Add 10%/°C of spec for offset and amplitude for operation outside the
range of 18° to 28°C.
2. Autorange enabled.
3. DC offset set to 0V.
4. Spurious output at low amplitude is –75dBm typical.
5. Add 1ppm/°C average for operation outside the range of 18° to 28°C.
6. FSK uses trigger input (1MHz maximum).
7. Sine and square waveforms above 10MHz are allowed only with an
“infinite” burst count.
BURST
7
WAVEFORMS: Sine, Square, Ramp, Triangle, Noise, ARB.
TYPE: Internal/External.
START/STOP PHASE: –360° to +360°.
INTERNAL PERIOD: 1µs ~ 500s.
GATED SOURCE: External trigger.
TRIGGER SOURCE: Internal, External, or Manual.
TRIGGER INPUT
LEVEL: TTL compatible.
SLOPE: Rising or falling (selectable).
PULSE WIDTH: >100ns.
IMPEDANCE: >10k W, DC coupled.
LATENCY: <500ns.
TRIGGER OUTPUT
LEVEL: TTL compatible into 1kW.
PULSE WIDTH: >400ns.
OUTPUT IMPEDANCE: 50W typical.
MAXIMUM RATE: 1MHz.
FAN-OUT: 4 Keithley 3390s.
PATTERN MODE
CLOCK MAXIMUM RATE: 50MHz.
OUTPUT: Level: TTL compatible into 2kW.
Output Impedance: 110W typical.
PATTERN LENGTH: 2 to 256K.
Model 3390 specifications
Model 3390 specifications