Instruction Manual

63
REMOVING AND EXCHANGING BLOCKS
The thermal cycling block can be removed and exchanged for one of a different format or for detailed
cleaning.
First ensure that the unit is disconnected from
the power supply and has been allowed to cool
if previously in use.
Open the drawer by pushing on the front
drawer panel then pull the drawer forwards to
access the block.
Remove the collar from the block by
unscrewing the four screws at the edges and
lift off the block.
Next, remove the orange insert behind the
draw front by unscrewing the two screws
beneath the panel, then lift out.
Slide the block forwards to release the
connectors and lift out of the unit.
Insert the new block and push back to connect
to the unit. Replace the orange insert and
screw in place.
Re-fit the appropriate collar for the block in use
– the standard collar for the 96 and 384 well
blocks or the deeper collar for the 60 x 0.5ml
block.