User Manual
TC-5000 SPECIFICATION
Temperature
Gradient Range 20°C to 70°C
Maximum Gradient 30°C
Minimum Gradient 1°C
Number of temperature probes 4
Block temperature range 4°C to 99°C
Typical heating rate 3.0°C/sec
Typical cooling rate 1.3°C/sec
Block uniformity (without gradient) <
+
_ 0.3°C
Column uniformity with a 10°C gradient <
+
_ 0.3°C
Control accuracy <
+
_ 0.3°C
Gradient accuracy <
+
_ 0.3°C
Temperature set point adjustment
+
_ 0.1°C
Heated Lid
The heated lid is only operational if the block temperature is set above 35°C
Selectable heated lid temperature 100°C to 115°C
Heated lid enable/disable Yes
Warm up time from ambient 4 minutes
Over-temperature cut-out Yes
Regulated heated lid pressure Yes
Programs
Password protected Yes
Incremental/decremental hold time Yes
Incremental/decremental temperature Yes
Variable programmable ramp rate steps
+
_ 0.1°C
Programmable ramp rate 0.1°C
Gradient calculator Yes
Maximum programs stored 50
Maximum number of segments per loop >60
Maximum number of loops per program 30
Program naming capacity Alpha numeric
Program interface Touch screen
Pause facility Yes
End of program alarm Yes (can be disabled)
Auto restart on power failure Yes
Run completion time Yes
21
Technical Specification at an ambient temperature of 20°C