SPECIFICATION
The DB-3D with RS232 Dri-Block
®
This unit has a recessed chamber which can hold up to three interchangeable
insert blocks which accept test tubes or other sample containers. The
Temperature is set by 3 front panel buttons with digital display.
The DB-3D is supplied with an extraction tool for removing the insert blocks
from the unit.
Working temperature range Ambient +5°C to 200°C
Settable temperature range 0.0°C to 200.0°C
Temperature stability ±0.1°C at 40°C
Temperature stability ±0.15°C at 100°C
Temperature setting 3 push buttons
Temperature display 4 digit LED
Set point resolution 0.1°C
Set point to accuracy ±1°C
Max temperature variation between blocks 0.2°C at 40° C
Heat up time 30° C to 37° C 11 minutes
Heat up time 30° C to 56° C 15 minutes
Heat up time 30°C to 200°C 25 minutes
Number of blocks 3
Electrical supply
Voltage Cycles Power
230V 50Hz-60Hz 650W
110V-120V 50Hz-60Hz 650W