The results table displays the calculated results.
Column headings:
No:
Well number.
Well ID:
Location of well.
Sample ID:
User-supplied or default name of well.
Cq (Dye 1):
Quantification cycle for Dye 1.
Conc (Dye 1):
Calculated concentration of sample for Dye 1.
Cq (Dye 2):
Quantification cycle for Dye 2.
Conc (Dye 2):
Calculated concentration of sample for Dye 2.
Ratio:
Dye 1 divided by Dye 2.
Comments:
User inputted text.
3.14.7.3 Relative quantification cycles
The relative Cq method allows the user to perform relative quantification without using a standard
curve. Instead, the Cq of an unknown sample is normalized to a known sample e.g. a reference
gene (REF) amplified at the same time from the same sample. For many gene expression studies
it is not necessary to determine the absolute amount of a target as evidence of a relative increase
or decrease in expression compared to a reference sample (calibrator) proves sufficient. The
expression level of the gene of interest in the calibrator sample is defined as 1x, with levels in the
unknowns reported relative to it.