User Manual

38
INSERT/DELETE A STAGE
A
stage
is a section of a program that consists of a number of individual steps that can be cycled
(repeated) a defined number of times.
Insert a new stage
To insert a stage in a program, touch
Insert a new
stage
. A stage with one step will be inserted.
Stage 1
Touch this to give the stage a name (optional).
Enter the name using the text entry screen.
1 cycle
Touch this to edit the number of cycles (repeats) for
the stage. Enter the required number of cycles then
touch
OK
.
Temperature
Touch the temperature button to edit the
temperature of the step and add any
increment/decrement or gradient features.
Note
: The minimum step temperature that can be
programmed during cycling stages is 10°C. If the step
temperature is above or below the valid gradient
range (PrimeG only) then the gradient function is
disabled.
Time
Touch the time button to edit the hold time of the
step and to add any increment/decrement time
features.
Ramp rate
Touch the ramp rate setting to adjust the ramp rate.
The range is 0.1ºC/s (minimum) to 3.4ºC/s maximum
for heating and 1.0ºC/s for cooling. The value can be
entered to 1 decimal place. Ramp rate will affect the
heating or cooling rate from the previous step to the
current step.
Once set, touch
OK
to return to the Programming
Screen.
Delete a stage
To
delete
a stage, touch the stage area to highlight
it and then touch
Delete
.