User Manual

19
DB200/2 AND DB200/3
Specification
DB200/2
DB200/2/115
DB200/3
DB200/3/115
Number of insert blocks 2 3
Temperature display 5 character LED 5 character LED
Working temperature range Ambient +5°C to 200°C Ambient +5°C to 200°C
Minimum set temperature C C
Set point resolution 0.1°C 0.1°C
Set point accuracy at 37°C ±1°C ±1°C
Set point accuracy at 100°C ±1°C ±1°C
Temperature stability at 37°C ±0.1°C ±0.1°C
Temperature stability at 10C ±0.15°C ±0.15°C
Max. temperature variation within a block
<0.2°C at 37°C
<0.2°C at 100°C
<0.2°C at 37°C
<0.2°C at 100°C
Max. temperature variation between
similar blocks
<0.5°C at 37°C
<1.5°C at 100°C
<0.5°C at 37°C
<1.5°C at 100°C
Heat up time 25°C to 100°C 8 minutes 7 minutes
Heat up time 25°C to 200°C 23 minutes 19 minutes
Countdown Timer with audible beeps Yes Yes
Minimum / Maximum timer 1min / 99h59min 1min / 99h59min
Safety over-temp cut-out Thermal fuse Thermal fuse
Overall dimensions (W x D x H) 202 x 260 x 105mm 279 x 260 x 105mm
Electrical supply 230V or 115V, 50-60Hz 230V or 115V, 50-60Hz
Power 300W 450W
Shipping weight 5kg 6kg
Note: The control specifications are quoted for an ambient temperature of 20°C for units fitted with
2, 3 or 4 insert blocks type F3506. The specification will be closely held if the ambient temperature is
in the range 10°C to 30°C. Outside this range the quoted figures may deteriorate but the unit will still
work safely.
Note: Different block inserts may take different amounts of time to equilibrate to the same
temperature.