SPECIFICATIONS
DB100/2, DB100/3 AND DB100/4
Specification
DB100/2
DB100/2/115
DB100/3
DB100/3/115
DB100/4
DB100/4/115
Number of insert blocks 2 3 4
Temperature display 5 character LED 5 character LED 5 character LED
Working temperature range
Ambient +5°C to
100°C
Ambient +5°C to
100°C
Ambient +5°C to
100°C
Minimum set temperature C C C
Set point resolution 0.1°C 0.1°C 0.1°C
Set point accuracy at 37°C ±1°C ±1°C ±1°C
Set point accuracy at 100°C ±1.5°C ±1.5°C ±1.5°C
Temperature stability at 37°C ±0.1°C ±0.1°C ±0.1°C
Temperature stability at 10C ±0.15°C ±0.15°C ±0.15°C
Max. temperature variation
within a block
<0.2°C at 37°C
<0.3°C at 100°C
<0.2°C at 37°C
<0.3°C at 100°C
<0.2°C at 37°C
<0.3°C at 100°C
Max. temperature variation
between similar blocks
<0.5°C at 37°C
<1.0°C at 100°C
<0.5°C at 37°C
<1.0°C at 100°C
<0.5°C at 37°C
<1.0°C at 100°C
Heat up time 25°C to 100°C 8 minutes 7 minutes 7 minutes
Countdown Timer with audible
beeps
Yes Yes Yes
Minimum / Maximum timer 1min / 99h59min 1min / 99h59min 1min / 99h59min
Safety over-temp cut-out Thermal fuse Thermal fuse Thermal fuse
Overall dimensions (W x D x H) 202 x 260 x 105mm 279 x 260 x 105mm 356 x 260 x 105mm
Electrical supply
230V or 115V,
50-60Hz
230V or 115V,
50-60Hz
230V or 115V,
50-60Hz
Power 300W 450W 600W
Shipping weight 5kg 6kg 7kg
Note: The control specifications are quoted for an ambient temperature of 20°C for units fitted with
2, 3 or 4 insert blocks type F3506. The specification will be closely held if the ambient temperature is
in the range 10°C to 30°C. Outside this range the quoted figures may deteriorate but the unit will still
work safely.
Note: Different block inserts may take different amounts of time to equilibrate to the same
temperature.