Instruction Manual

ADDITIONAL INFORMATION
ACCESSORIES
Insert blocks are made of aluminium alloy and must be ordered separately from the heater units. The
following blocks are available from Techne:
Part Tube Size Number of Holes Size
F3501 Plain block None 95 x 76 x 51
F3502 6mm 30 95 x 76 x 51
F3503 10mm 20 95 x 76 x 51
F3504 12mm 20 95 x 76 x 51
F3505 13mm 20 95 x 76 x 51
F3506 15mm 12 95 x 76 x 51
F3507 16mm 12 95 x 76 x 51
F3508 19mm 8 95 x 76 x 51
F3509 25mm 6 95 x 76 x 51
F3510 10mm cuvettes 2 channels 95 x 76 x 51
F4460 Plain block Thermometer hole only 95 x 76 x 51
F4461 9mm/7mm 10/20 95 x 76 x 51
F4462 24mm 6 95 x 76 x 51
F4463 26mm 6 95 x 76 x 51
F4464 1.5ml Eppendorf 20 95 x 76 x 51
F4465 0.5ml 30 95 x 76 x 51
F4466 Plastic half block Plain 95 x 37 x 50
F4470 2 ml Eppendorf 20 95 x 76 x 51
F4471 0.2ml 72 95 x 76 x 51
1. Overtemperature cutout, DB-3D and DB-2TC
The overtemperature cut-out is a sensitive mechanical device and mechanical shock can cause it to trip.
• In the event of no heater power, check the mains plug and lead, then reset the cut-out control (located
at the rear of the unit).
• Repeated operation of the cut-out indicates a serious fault: you may need to return the unit to your
supplier for repair.
2. Fuses
Your unit is protected by two fuses. These should only be changed by suitably qualified personnel. If the
fuses blow persistently, a serious fault is indicated and you may need to return the unit to your supplier
for repair.
DB-2D 230V 2 x F2A; 120/100V 2 x F4A
DB-3D 230V 2 x F4A; 120/100V 2 x F6.3A
DB-2TC 230V 2 x F2A; 120/100V 2 x F4A
28