Operation Manual
c) Wrap the cuff around your upper left arm.
The blue arrow (red mark) should be on the inside of your arm extending downward to your
hand. Make certain the cuff lies
approx. 1–2cm above the elbow.
Important! The red mark on the
edge of the cuff (Artery Mark) must
lie over the artery which runs down
the inner side of the arm.
d) There should be little free space between
your arm and the cuff. You should be able to fit
two fingers between your arm and the cuff. Cuffs
that don’t fit properly result in false
measurement values. Measure your arm
circumference if you are not sure of proper fit.
The arrow in the picture marks the artery.
Place the middle of the CUFF at the level of the
right atrium of the heart
e) Lay your arm on a table (palm upward) so the cuff is at the same height as your heart.
f) Remain seated quietly for at least two minutes before you begin the measurement.
5.4. Measurement procedure
General recommendation: approx. 5 minutes should elapse before the first reading is
taken. The performance of the AUTOMATED SPHYGMOMANOMETER can be affected by
extremes of temperature, humidity and altitude.
a) Sit comfortably in a chair with your feet flat on the floor (no crossed legs) and your back
and arm are supported.
b) Stretch your arm forward on the desk and keep relaxing, make sure the palm of hand is
upturned. Make sure arm is in correct position, to avoid body movement. Sit still and do not
talk or move during the measurement. After the cuff has been appropriately positioned on
the arm and connected to the blood pressure monitor, the measurement can begin.
(See now chapter 6. Operate on the device via APP)
c) To stop the inflation or measurement: If it is necessary to interrupt a blood pressure
measurement for any reason (e.g the patient feels unwell), press the START/STOP button.
The monitor will stop inflating, start deflating, and will turn off. If the monitor is remote
operated by APP on smart phone, it also can be stop on APP.