Data Sheet

Specication (See www.uk.tdk-lambda.com/zplus for further information)
Z+ Series
2
Notes:
(1) Recovery to within 0.5% of rated voltage after a load change of 10-90% (Output current 10-100% of Imax)
(2) During 30 minute warm up time after power on
(3) Requires IEC 320-C15 connector for mains input
Model Z10 Z20 Z36 Z60 Z100
Load Regulation CV 2mV + 0.01% of rated voltage over 0 - 100% load change
Line Regulation CV 2mV + 0.01% of rated voltage over a 85 - 132 or 170 - 265VAC line change
Recovery Time (1) CV 1ms
Temperature Coefcient CV 30ppm/C following 30 minute warm up
Temperature Stability CV 0.02% of rated voltage over 8 hours following 30 minute warm up time
Warm up Drift (2) CV <0.05% of rated voltage + 2mV of rated output voltage
Up programming response time
(10-90% or 90-10% of Vmax)
CV 15ms 30ms 30ms 50ms 50ms
Down programming resp time
(CV) (10-90% or 90-10% of Vmax)
Full load 10ms 30ms 30ms 50ms 50ms
Down programming resp time
(CV) (90-10% of Vmax)
Zero load 190ms 200ms 250ms 310ms 900ms
Load Regulation CC 5mA + 0.01% of rated current over 0 - 100% Vout change
Load Regulation thermal drift CC < 0.05% of rated current over 30 minutes after load change
Line Regulation CC 2mA + 0.01% of rated current over a 85 - 132 or 170 - 265VAC line change
Temperature Coefcient CC 100ppm/C of rated current after 30 minute warm up time
Temperature Stability CC 0.05% of rated current over 8 hours following 30 minute warm up time
Warm up Drift (2) CC <+/-0.1% of rated current
Vout & Iout programming & readback resolution Digitally < 0.012% of rated voltage/current
Vout & Iout programming & readback accuracy Digitally < 0.05% of rated voltage, < 0.1% of rated current
Voltage & Current Programming Analog By either Voltage (0-5V or 0-10V) or Resistance (0-5k or 0-10k)
Voltage & Current Monitoring Analog 0-5V or 0-10V Voltage (user selectable), +/-1% accuracy
Overvoltage Shutdown (user programmable) V 0.5 - 12 1 - 24 2 - 40 5 - 66 5 - 110
Overtemperature Protection - User selectable - latched or non-latching
Display - Voltage - 4 digits. Accuracy 0.5% of rated voltage or current +/- 1 count
Remote On/Off - By applied voltage or dry contact relay (user selectable logic)
Output Good - Open Collector, Low on fail
Remote Sense Compensation (per wire) V 1 1 2 3 5
Communication Interface - RS232, RS485 & USB standard, IEEE488 (GPIB) & LAN optional
Series Operation - Up to two identical units (with external diodes)
Parallel Operation - Up to ve units in master-slave conguration
Input Voltage / Frequency (3) VAC 85-265VAC, 47-63Hz
Inrush Current A < 25A
Hold Up Time (Typical) ms 16ms
Power Factor Correction - Complies with EN61000-3-2 Class A (0.99 typ)
Operating Temperature °C 0 - 50°C
Storage Temperature °C -20 to +85°C
Humidity (non condensing) %RH Operating: 10 - 90%RH, Storage 10 - 95%RH
Cooling - Internal temperature controlled fan
Withstand Voltage VAC I/P to GND 2kVAC, I/P to O/P 3kVAC, O/P to GND 1380VDC 1 min
Insulation Resistance Ω >100MΩ at 25°C & 70%RH
Vibration (non operating) - IEC60068-2-64
Shock G <20G, half sine, 11ms. IEC60068-2-27
Safety Agency Certications - UL61010-1, EN61010-1, IEC61010 (Designed to meet UL/EN60950-1)
Immunity - IEC61326 (Designed to meet EN55022 / EN55024)
Conducted & Radiated EMI - EN55022-B, FCC part 15-B, VCCI-B
Size (H x W x D) (Excluding handles and busbars) mm Standard body 83 x 70 x 350mm; Wide Body 83 x 105 x 350mm
Weight kg 200W & 400W: Standard body 1.9kg, Wide body 2.4kg
600W & 800W: Standard body 2.5kg, Wide body 3.0kg
Warranty yrs 5