Data Sheet
HWS30-1500/ME Series
1
HWS30-1500/ME Series
Single Output
Medical Power Supplies
Key Market Segments & Applications
Non-Surgical Medical Equipment
Analysers
•
Limited Lifetime Warranty
•
Medical Approvals
•
Universal Input (85 - 265VAC)
•
High Efficiency
•
Broad 30W to 1500W product range
•
RoHS Compliant Design
HWS30-1500/ME Features and Benefits
Features Benefits
• Limited Lifetime Warranty • Lower Cost of Ownership
• Medical Approvals • Reduces System Approval Times
• Wide Range AC Input • Supports Global Use
Specifications
Notes: (1) 100/200VAC input (2) See clause 19.5DV.2 of UL60601 for equipment in proximity of patient (3) HWS300-600 2.5kVAC Input to ground
MODELS HWS30 HWS100 HWS300 HWS600 HWS1000 HWS1500
ITEM HWS50 HWS150
Input Voltage range
85-265VAC (47-63Hz) or 120-370VDC 85-265VAC (47-63Hz) or 120-330VDC 85-265VAC (47-63Hz)
0.8 / 0.4 1.3 / 0.65
Input Current (Typ) (1) A 0.7 / 0.35 1.9 / 0.95 4.1 / 2.1 8.1 / 3.9 13.5/7.0 19/10
Inrush Current (1) A 14 / 28 20 / 40
Power Factor / Flicker Meets EN61000-3-2, EN61000-3-3
Temperature Coefficient - <0.02%/°C
Overcurrent Protection - >104%
Overvoltage Protection V Yes (See table on page 2)
Hold Up Time (Typ) ms 20
Leakage Current (60Hz) (2) mA <0.5mA
Remote Sense No Yes
Indicator - Green LED = ON
Remote On/Off - No Yes (Isolated from output)
Parallel operation - No Single wire conn. (5 units max)
DC Good - No Yes
Voltage Programming No Yes
Operating Temperature - HWS30-150: -10°C to +70°C, (-10 to +50°C: 100%, +60°C: 60%, +70°C: 20%)
and Derating HWS300-1500: -10°C to +70°C, (-10 to +50°C: 100%, derate linearly to 50% load from +50°C to +70°C)
Storage Temperature °C -30°C to +85°C
Humidity (non condensing) - Operating: 30 - 90%RH (10 -90% on HWS300-1500), Non operating 10 - 95%RH
Cooling - Convection Internal fan
Withstand Voltage (3) - Input to Ground 2kVAC, Input to Output 3kVAC, Output to Ground 500VAC for 1 min.
Isolation Resistance - >100MΩ at 25°C & 70%RH, Output to Ground 500VDC, >10MΩ Output to remote on/off 100VDC
Vibration (non operating) - 10 - 55Hz (1 min sweep), 19.6m/s
2
constant, X, Y, Z axis, one hour each
Shock (in packaging) - < 196.1m/s
2
Safety Agency Approvals (2) - UL60601-1, EN60601-1, CSA-C22.2 No6011-M90 (C-UL) (basic insulation), CE Mark
Line Dip - Complies with SEMI F47 (200VAC line only)
Conducted & Radiated EMI - EN55011 / EN55022-B, FCC-B, VCCI-B (HWS600 & 1500 Class A)
Immunity - IEC61000-4-2 (Level 2,3), -3, -4, -6, (Level 3), -5 (Level 3,4), -11
220 450
Weight (Typ) g 280 500 1000 1600 3200 3800
26 x 82 x 95 28 x 82 x 160
Size (WxHxD) mm 26 x 82 x 120 37 x 82 x 160 61 x 82 x 165 100 x 82 x 165 126.5 x 82 x 240 280 x 82 x 126.5
Warranty Limited Lifetime Warranty