Datasheet

CUS30M & CUS60M Series
,QGXVWULDO 7HVW
Single Output 30 to 60W Medical &
ITE Power Supplies
CUS30M & CUS60M Series
1
www.emea.tdk-lambda.com/cus30
_
60m
&RPP
%URDGFDVW
Features Benets
•HighEfciency,upto90% •EasierThermalManagement
•Compact2”x3”Footprint •SpaceSaving
•ConvectionCooled •QuietOperation
•SuitableforClassIandIIInstallations •FlexibleUtilisation
•MedicalApprovalwith2xMoPPIsolation •SuitableforBandBFRatedEquipment
Specication
Model CUS30M CUS60M
ACInputVoltage VAC 85-265VAC
(1)
InputFrequency Hz 47-63Hz
InrushCurrent(coldstart) A 30Aat115VAC,60Aat230VAC
PowerFactor - MeetsEN61000-3-2
InputCurrent(115/230VAC) A 0.6/0.4 5V:0.7/0.5,12-48V:1.2/0.8
Off-loadPowerDraw W <0.3Wat265VAC <0.5Wat265VAC
TemperatureCoefcient %/°C <0.02%/°C
OvercurrentProtection - >105%ofratedcurrent,Class2LimitedPowerSource
(2)
OvervoltageProtection
(3)
V >115%ofnominaloutput >120%ofnominaloutput
HoldUpTime(Typical)(115/230VAC) ms 20/100
LeakageCurrent mA <0.25mAat265VAC,60Hz <0.2mAat265VAC,60Hz
TouchCurrent(Maximum) µA 100µAat265VAC,60Hz 60µAat265VAC,60Hz
RemoteSense - No
OperatingTemperature °C -20°Cto+70°C.(-30°Cstartup)Deratelinearlyto50%loadfrom+50°Cto+70°C
(4)
StorageTemperature °C -40°Cto+85°C
Humidity(noncondensing) %RH 10-95%RH(Operating&Storage)
Cooling - ConvectionCooled
WithstandVoltage -
InputtoGround2kVAC(1xMOPP),InputtoOutput4kVAC(2xMOPPs),OutputtoGround
1.5kVAC(1xMOPP)suitableforB&BFratedequipment
IsolationResistance MΩ >100MΩat25°C&70%RH,OutputtoGround500VDC
Vibration(nonoperating) - 10-55Hz:19.6m/s
2
constantsweep1minX,Y,Zfor1hour
Shock - <196.1m/s
2
,MIL-STD-810F
Immunity - IEC61000-4-2(lv4),-3(lv3),-4(lv3),-5(lv3,4),-6(lv3),-8(lv4),-11,CriteriaA.IEC60601-1-2Ed4
LineVoltageDips -
SEMIF47(200-240VAC),IEC61000-4-11(Class3)
CriteriaA:Vin>120VAC,or<70%load,CriteriaB:Vin<120VAC,and>70%load
SafetyAgencyCertications - EN/IEC/UL/ES/CSA60601-1,EN/IEC/UL/CSA60950-1,CEMark,IEC60335-1
Altitude m 5,000m,derateoperatingambientby5°C/1000mabove3,000m
Conducted&RadiatedEMI - EN55011-B,EN55032-B,FCCClassB
Weight(Typ) g 62 120
Size(LxWxH) mm 76.2x50.8x24.2 76.2x50.8x26.7
Warranty yrs 3
0HGLFDO
Notes:
(1)CUS30M:Deratelinearlyto80%loadfrom115VACto85VAC.CUS60M:Deratelinearlyto80%loadfrom100VACto85VAC
(2) AllmodelsexceptCUS60M-5
(3)CycleACtoreset
(4)CUS30M-18and-48,CUS60M-5and-12:Deratelinearlyto50%loadfrom+45to+70°C