3. Description
The measurement modes analogue signal 0/2 to 10 V and 0/4 to 20 mA DC are
supported.   The   multi   display   is   able   to   display   two   parameters   simultaneously.
Thresholds   can   be   associated   with   individual   warning   colours.   Minimum   and
maximum values  are  recorded  and  can  be   displayed and reset  via jumper  or  an
external switch. Measured values of one parameter are recorded over a timespan of 3
minutes up to 14 days. The timebase as well as the display of the measuring history
can be activated by an external switch. The measured values remain stored as long
as the device is supplied with voltage.
The  following  parameters   can   be   adjusted  using   the   USB  configuration   software:
scale   endpoint,   scale   caption,   display   style   (pointer,   tachometer,   bar   graph,   and
more), splash image, backlight colour, thresholds for alarm output, threshold warning
colour (light, blink), hysteresis, and more. Recorded measurement values can be read
and a continuous transmission of the actual measuring value can be enabled. Using
the software, values can be viewed and exported as CSV.
Front
1 Unit or free text
2 Threshold values
3 Upper scale caption
4 Graphical measurement display
5 Digital measurement value
6 Lower scale caption
Back
1 2 optocoupler switch outputs
2 DIP switch for setting measurement
mode and upper scale caption
3 Supply voltage input
4 Jumper pins J1 – J6 for 
activating/switching of 
measurement mode, graphical 
historic data display, min/max 
display, auto-scaling
5 USB interface for alternative power 
supply and software communication
interface
6 Measurement inputs
----- 2 -----