Instructions
Table Of Contents
9. Other settings
The following functions can be activated independently during operation by short-
circuiting connectors J1-J3, e.g. using a jumper or switch:
J1: Graphical historic data display
The   unit   displays   the  stored   values  within
the set time base as a graphical trend. The
time   base   can   be   set   to   days   (7,   14),
hours (1, 3, 6, 12, 24, 48, 72), minutes (3,
15, 30) or seconds (36).
The   time   base   can   be   changed   by
alternately opening and closing J1 (interval
< 2 sec). When first opening and closing J1
the current time base is displayed. For each
subsequent   opening   and   closing   the   time
base changes to the next setting. In order to
save the setting  permanently,  the  software
“Digalox Manager” has to be used.
J2: Min-Max display
The   display   shows   the   maximum   and
minimum   values   recorded   since   the   last
reset. The values can be reset by opening
and closing connection J2 shortly (interval <
2 sec). The display shows “Minmax reset”.
J3: Auto-scaling
The device automatically changes the upper
scale   caption   depending   on   the   current
measuring value between 10, 100 and the
set upper scale caption.
----- 6 / 8 -----