Instructions
----- 3 -----
4. Electrical Connections
The device may only be operated in one of the connection options shown below.
Warning:
Voltage measurement on the high
side (between plus and load) is
recommended only for voltages up
to 50V!
In this case the entire device is on
high potential. Particularly to be
considered when connecting
switches to JP1-6.
DPM72 AV/P
Voltage and current
direct measurement
I
low
I
high
U
high
U
low
U
s
U
s
Load
+/L -/N
max.
10A
Measurement inputs
10-24V
Supply
Fuse 250mA
max.
500V
DPM72 AVP only!
Voltage and current direct
measurement, high side
I
low
I
high
U
high
U
low
Load
+/L -/N
Measurement inputs
AVP
only!
max.
50V!
U
s
U
s
10-24V
Supply
Fuse 250mA
DPM72 AV/P
Voltage direct, current via
current transformer
I
low
I
high
U
high
U
low
Load
+/L -/N
Current
transformer
Measurement inputs
max.
500V
U
s
U
s
10-24V
Supply
Fuse 250mA