User's Manual
30
3.After making all the desired settings, click the [Start Recording]
button to transmit them.
4.After transmission has been completed the communication result
will appear. By clicking [OK], the set-up process will be nished. By
clicking [Close], you can close the window.
■
[
Get Settings
]
Button
The Remote Unit recording settings, registration contents, and remaining
battery life are displayed.
●
Registration contents: If a settings transmission was received it displays
the Group and Remote Name(s).
●
Remaining battery life: If a settings transmission was received it displays
the remaining battery life.
Voltage/Average Value: Record as the average value for each recording
interval span. The average value for recording
intervals under 15 seconds will be calculated as
the average of the measurements from every 1
second. The average for intervals over 20
seconds will be calculated as the average of
the measurements from every 2 seconds.
Pulse/ Lo→Hi: Record the number of signals rising within the
recording interval you have set.
Pulse/ Hi→Lo: Record the number of signals falling within the
recording interval you have set.
Pulse/ Event: Record the time of any event; a rising (Lo to Hi)
or falling (Hi to Lo) waveform that occurs for
more than 1 second at an input voltage range of
between 0-30V.
※
By clicking the [Get Settings] button, registration contents and remaining
battery life will be displayed.