User's Manual
Ķı
Adjustment Function
This function enables the user to make adjustments to measured values to correct
inaccuracies that may occur when compared to a standard or reference temperature
measurement. By setting an adjustment equation beforehand, the adjustment function will
automatically calculate to record and display the post-adjusted measurement. It is possible
to select from two methods of adjustment: 1-point adjustment and 2-point adjustment.
Note:
When you are carrying out adjustments and formulating equations please take care with the settings and take
personal responsibility for the process.
ɡ 1-point / 2-point Adjustment
Before : Enter the actual measurement before adjustment
After : Enter the desired measurement after adjustment
- Entries are valid to one decimal point.
- The input value for [Post-adjusted Reading] must be less than Ⱦ 5% in comparison to the
input value for [Current Reading].
ɡ Adjustment Settings
1.
Open [RTR-5W for Windows].
2.
In the Main Window, click on the Data Logger icon. To open the [Adjustment] Window,
click on [Settings] in the Menu Bar and then select [Adjustment Settings] from the pull
down menu.
<RTR-51, 52, 53>
[Send Settings]
button
[Initialize]
button
<RVR-52A>
[Send Settings]
button
[Initialize]
button
*
3.
Select either [1-point Adjustment] or [2-point Adjustment].
4.
Please enter the actual pre-adjusted value(s) in the [Current Reading] column and the
post-adjusted reference value(s) in the [Post-adjusted Reading] column.
5.
Click on the [Send Settings] button to send the adjustment equation to the selected
Data Logger.