User’s Manual Thank you for purchasing our product. Carefully read this instruction manual before using this unit. Copyright 2001-2003 T&D Corporation. All rights reserved. 2004.
■ Notices about this User’s Manual In order to properly use this product, please carefully read this manual before using.T&D Corporation accepts no responsibility for any malfunction of and / or trouble with this product or with your computer that is caused by the improper handling of this product and will deem such trouble or malfunction as falling outside the conditions for free repair outlined in the attached warranty. ● All rights of this User’s Manual belong to T&D Corporation.
Safety Precautions and Instructions To ensure safety be sure to obey all of the following warnings. The following items should be strictly obeyed for the safe usage of this unit, and for protecting yourself and other people from bodily harm and/or damage to property. To ensure the proper use of our product, please read the following carefully and fully understand the contents.
DANGERS Be sure to follow the warnings and notices about use from your PC maker when installing and using this unit. Do not take apart, repair or modify the main unit. Doing so may cause fire or electrocution. If water or a foreign body enters in this unit, turn the power off and remove the batteries. Continued use may cause fire or electrocution. Do not use this unit in wet or humid places, such as a bathroom. Store RTR-57C main units, batteries and communication cables out of the reach of children.
CAUTIONS This unit is not water-resistant. If the unit gets dirty, wipe it with a clean cloth and a mild detergent. Do not put your fingers or foreign matter into the communication port. Battery life depends on the kind of the batteries, measurement environment, communication frequency, ambient temperature, and battery quality. Do not use batteries other than specified. Doing so may cause fire or damage. Do not use an AC adaptor other than specified. Doing so may lead to fire or damage.
CAUTIONS Do not expose the unit to harmful gases or chemicals. It may cause corrosion and/or other danger to the unit and to people handling the unit. To prevent damage to the unit from static electricity, remove static electricity from your body by touching metal around you (door knob, window frame) before touching the unit. Static electricity may cause not only damage to the unit, but may cause breaks in or a loss of data. Do not use the unit in places such as listed below.
■ FCC Compliance Statement for American Users This device complies with Part 15 of the FCC Rules. Operation is subject to following two conditions: (1) this device may not cause harmful interference. and (2) this device must accept any interference received, including interference that may cause undesired operation. Note: This equipment has been tested and found to comply with the limits for a Class A Digital Device, pursuant to Part 15 of the FCC Rules.
Table of Contents Introduction Safety Precautions and Instructions ...................... ⅱ About Data Collector RTR-57C .............................. 1 Contents of Package .............................................. 4 Part Names and Functions ..................................... 5 Getting Ready Installing the Batteries ........................................... 9 Switching the Power on ......................................... 10 Setting the Date and Time .....................................
Other Functions Checking Upper and Lower Limits ・Setting the Upper and Lower Limit Range ...... 41 ・Viewing the Results of Saved Data................. 42 Remote Unit Conditions Display............................ 43 Searching for Remote Unit .................................... 45 Monitoring.............................................................. 47 Displaying Recorded Data in Graph ...................... 49 Deleting Data ・Deleting Selected Data ...................................
About Data Collector RTR-57C ◆ Outline Our revolutionary Wireless RTR-5 series is a system wherein the various kinds of data measured and recorded by our data logger units (RTR-51/RTR-52-Temperature data, RTR-53-Temperature and Humidity data, and RVR-52-Voltage, Pulse and Event data) can be transmitted to the RTR-57C Data Collector via short wave radio communication. The Data Collector can then be connected to your computer to enable easy downloading and total data management.
● Easy Data Collection from a Variety of Unit Types Data downloading can be carried out with any RTR-5 Series Units via wireless communication or by simply placing it face down on the communication pad area. D a t a f r o m o t h e r t y p e s o f u n i t s c a n b e d ow n l o a d e d v i a t h e p r ov i d e d communication cable. ● Data Capacity: 260,000 Readings The RTR-57C has a large enough capacity to collect data from 16 RTR-51 units at full capacity.
● Battery Life Warning Display When the battery power goes low, a battery life warning will be displayed. If the battery is not changed the unit will automatically go into sleep mode. ● Trustworthy Backup Function If, after the battery warning display appears, the battery power goes even lower the unit will automatically go into sleep mode. Although, in sleep mode, the unit cannot be operated and power cannot be turned on, all data will be safely stored without loss.
Contents of Package The following items are included in the package.
Part Names and Functions ◆ Part Names ⑦ ① ② ⑤ ⑥ ③ Collect List ④ ① Operation Dial ② LCD Display ③ Operation Buttons ④ Optical Communication Port ⑧ ⑤ Communication cable connection jack ⑥ AC adaptor connection jack ⑦ Antenna ⑧ Battery Compartment ◆ Explanation of the Button Functions ・When downloading via wireless communication the [Select Group] display will appear. ・Allows you to swift-scroll across data when viewing a graph or an event list.
◆ How to Use the Operation Dial Move the dial up Press in Move the dial down MAIN MENU GATHER DATA SEARCH RMOTE MONITOR MODE GATHER DATA DATA LIST REC SETTINGS 〔Moving the dial up and down〕 ● By moving the dial up or down the arrow will move to allow you to choose the desired item. The selected part will be displayed in an easy-to-see inversed shade. ● When setting a numerical value, by moving the dial up the value will become larger and by moving the dial down the value will become smaller.
◆ LCD Display marked with( )indicate menus for Wireless Communication. ● Items item marked with an arrow( )denotes it has been selected. If you move the operation dial up and down, the arrow will move accordingly and by pressing in the dial the item selected will be activated. ● The ●A▲ indicates that there are more menu items above ( ▲ ) or below the ( ▼ ) to view. ● After making a setting, adjust the dial to select [OK] and press to complete the setting procedure.
● If there is no [OK] in the setting display, but a [BACK] is displayed, adjust the arrow to [BACK] and press it to complete the setting and take you back to the menu display. ● When carrying out a Remote Unit search, a mark will appear next to Group Names in which Remote Units that can carry out wireless communication are registered, as well as, next to any Remote Unit name that can carry out wireless communication.
Installing the Batteries ◆ Install 2 AAA alkaline batteries. ● AAA Ni-Cd and AAA Ni-MH(1.2V) batteries sold on the market can also be used. NOTE: ● Install 2 of the same kind of batteries. ● Install the batteries in the correct direction to align + and -. ● Battery charging via the main unit is impossible. ● If you use an outside power source, batteries are not necessary. ● The RTR-57C unit has no battery recharging ability.
Switching the Power on ◆ Switch the power ON by pressing the POWER button or the Operation Dial. 《Press the POWER button》 《Press the Operation Dial》 ● To switch the power OFF, press the [POWER] button. You cannot switch the power OFF with the Operation Dial. ● During communication, even by pressing the [POWER] button, it is impossible to switch the power OFF.
Setting the Date and Time CAUTION ! ● After purchasing this product, please be sure to use the attached software to set the clock and date. ● The difference between your local time and GMT will automatically be calculated and written, after which the correct time will be displayed.
Communicating with the Data Logger There are two ways to communicate between the RTR-57C and the Data Logger. 【Via Cable / Optical Communication】 Via Communication Cable connection for TR-7 Series and Optical Communication for RTR-51/52/53, RVR-52 and TR-5 Series. 〔For RTR-51/52/53,RVR-52 and TR-51A/52〕 Place logger face down on RTR-57C 〔For TR-71S/72S〕 Connect using the included communication cable (TR-3C10) ◆ The following operations can be done using communication via cable or optical communication.
【Wireless Communication】 Communication can occur via special shor t radio waves with RTR-51/52/53 and RVR-52. Group 100 meters Remote Unit Remote Unit ● Transmission distance is 100 meters if direct and unobstructed. NOTE: Wireless communication can only occur with those RTR-51/52/53 and RVR-52 logger units that have been registered as Remote Units via the accompanying software『T&D Recorder for Windows 』. ◆ The following operations can be done using wireless communication.
● Read Logger Settings The RTR-57C can read and display all settings for any selected data logger that has been registered. ● Search for Remote Unit (s) A mark appears next to the name of any Remote Unit with which wireless communication is possible. ● Monitor Current Readings You can monitor the current measurements of a Remote Unit at a set interval or in groups with the measurements displayed consecutively. ◆ One Remote Unit can be set up to communicate with any number of RTR-57C Units.
Communicating with the Computer To download collected data to your computer, connect the RTR-57C Unit to your computer with the communication cable provided. ※ Downloading, editing and managing collected data can be done with the provided software. For details about how to use the software, refer to the User’ s Manual that comes with『T&D Recorder for Windows 』or the [Help Menu] in 『T&D Recorder for Windows』. NOTE: Wireless communication cannot be done with the computer.
2.Connect the communication cable to the RTR-57C. ● To prevent faulty contact, make sure to fully insert the communication cable.
【Communication between RTR-51/52/53 RVR-52 and your Computer】 1.Connect the communication cable. ① Connect the RTR-57C Unit to your computer with the provided communication cable (RS-232C). ② Place the data logger on the RTR-57C. ゲ ー アク ュ イサ ニ スン ト メ タ ケ ング ス ゲリ ー ン デ キ リア ・ イ L コ タイ サ メ W L モニ ス ウ シ ソ WL タ カイ Wー タ デーク デロ キ Place the unit face down on the RTR-57C. Communication Cable (to computer) NOTE: If light enters during communication, an error will occur.
Wireless Communication: Remote Unit Registration NOTE: ● Wireless communication can only be carried out with RTR-51/52/53 and RVR-52 units that have been registered as Remote Units via the provided software『T&D Recorder for Windows』. ● Please extend the antenna when using wireless communication. ◆ Register the data logger unit to be used as a Remote Unit. 1.Connect the cable so that communication can occur between the data logger and your computer. (See p.
Wireless Communication: Recording Start The following settings and changes can be made from the main unit: ● Recording Start Date and Time (Programmed Start) Mode ● Recording Interval ● Recording C or ゜F) in the RTR-51/52/53 units and the ※ To change the unit of temperature ( ゜ measurement mode in the RVR-52 units, please use your computer and the software『T&D Recorder for Windows 』. NOTE: ● If you make recording condition changes while recording, all previously recorded data will be lost.
● Before selecting a Remote Unit you can run a search to find all Remote Units with which communication is possible. Those units with which communication is possible will appear with a mark next to them. (See p.45 for details) 3.Make Recording Condition Settings. ● If you wish to make no changes and want to restart recording under the same conditions, please read the current settings first and then restart recording. (See p.
5.Once transmission has been completed, a message will appear on the display telling you communication was completed successfully. By pressing the operation dial you can go back to the Recording Settings Menu Display. The Remote Unit(s) will begin recording.
【Programmed Start】 1.In the Main Menu, click [REC SETTINGS] and then [ Wireless]. 2.Select the Remote Unit for which you wish to make settings. ① Select a Group in which Remote Units are registered. ② Select a Remote Unit or a Group. ● Before selecting a Remote Unit you can run a search to find all Remote Units with which communication is possible. Those units with which communication is possible will appear with a mark next to them. (See p.45 for details) 3.Make Recording Condition Settings.
4.After having completed the settings, adjust the arrow (>) to [REC Programmed] and press dial. 5.After having set the desired recording start date and time, adjust the arrow (>) to [REC SETTINGS] and press the dial to transmit the info. 6.Once transmission has been completed, a message will appear on the display telling you communication was completed successfully. By pressing the operation dial you can go back to the Recording Settings Menu Display. The Remote Unit(s) will be put on Record Standby.
◆ About Forbidding Recording Start by Wireless ● If a selected Remote Unit has been set to only start recording via computer and has been to set to be protected against changes in the recording start via wireless communication, the following message will appear. ● When starting recording by Group, a list will appear showing the Remote Units with which communication occurred.
Wireless Communication: Downloading Recorded Data About 2000 data readings can be downloaded per minute. ※ About 1000 event data readings can be downloaded per minute. NOTE: ● Wireless communication can only be done with RTR-51/52/53 and RVR-52 units that have been registered as Remote Units via the software『T&D Recorder for Windows』. ● Please extend the antenna before carrying out wireless communication. 1.In the Main Menu, open [ GATHER DATA]. 2.Select a Remote Unit.
3.Click [Download Data]. 4.Set the period of data you wish to download and click [START GATHER] to start downloading. 【Period of Data】 You can set the period (how many hours or days before now) from which you wish to download data. ● You can select from 1 ~ 47 hours (in 1 hour units) / 2 ~ 300 days (Less than 100 days in units of 1 day, More than 100 days in units of 5 days)/ or ALL DATA.
5. When the downloading is completed, a message will be displayed. By pressing the operation dial, the display will change to that of a graph. ● If you have set the judgment [ON] in [Set ↑↓ Limits] before downloading, the result will be displayed when the downloading is completed.
◆ If the amount of main unit memory is low, the downloading of recorded data may be impossible. When you check the amount of the memory to confirm whether the data you wish to download can be saved in the RTR-57C, and the data cannot be saved, the message [Memory Full] will be displayed and the downloading will be canceled.
Cable Communication: Recording Start The following settings or changes can be made via the main unit. ● Recording Start Time Setting (Programmed Start) Mode ● Recording Interval ● Recording ※ Please make temperature unit changes for the RTR-51/52/53 main units and Measurement Mode changes for the RVR-52 units via the supplied software『T&D Recorder for Windows 』 NOTE: ● If you set the conditions during recording, all previously recorded data will be lost.
Recording Modes: One Time: Recording will stop when the data capacity becomes full. Endless:Recording will continue with oldest data continuously overwritten. Recording Interval: Choose from 15 intervals: 1,2,5,10,15,20,30 seconds and 1,2,5,10,15,20,30 and 60 minutes. ※ Some recording intervals cannot be set on some types of the devices. 3.Move the arrow to [Immediate] after setting the conditions. 4.Connect the data logger to the RTR-57C and press the operation dial to have the settings transmitted.
5.When communication has been completed the message [COMM END] will be displayed and by pressing the operation dial, the display will return to the menu. The Remote Unit will star t recording. ◆ About the Default Settings At anytime, you can return the RTR-57C to its default settings. Recording Mode: Endless Recording Interval: 10 minutes Channel Names: 1 channel…ch.1, 2 channel…ch.
【Programmed Start】 1.In the Main Menu, open [REC SETTINGS]-[Direct]. 2.Set the recording conditions. Recording Modes: One Time: Recording will stop when the data capacity becomes full. Endless: Recording will continue with oldest data continuously overwritten. Recording Interval: Choose from 15 intervals: 1,2,5,10,15,20,30 seconds and 1,2,5,10,15,20,30 and 60 minutes. 3.Move the arrow to [Programmed] after setting the conditions. 4.Set the recording start time and move the arrow to [REC START].
5.Connect the data logger to the RTR-57C and press the operation dial to have the settings transmitted. 〔For RTR-51/52/53,RVR-52 and TR-51A/52〕 Place logger face down on RTR-57C ● If light enters during communication, an error will occur. Please make sure to place the unit within the ridges for an exact fit. 〔For TR-71S/72S〕 Connect using the included communication cable (TR-3C10) 6.
Cable Communication: Download Recorded Data Download recorded data by connecting the data logger to the RTR-57C. 1.Connect the data logger to the RTR-57C. 〔For RTR-51/52/53,RVR-52 and TR-51A/52〕 Place logger face down on RTR-57C ● If light enters during communication, an error will occur. Please make sure to place the unit within the ridges for an exact fit. 〔For TR-71S/72S〕 Connect using the included communication cable (TR-3C10) 2.In the Main Menu, open [GATHER DATA] and downloading will start.
● With RTR-51/52/53 and RVR-52 you can select to download data from any number of hours or days before now. ・You can select from 1 ~ 47 hours (in 1 hour units) / 2 ~ 300 days (Less than 100 days in units of 1 day, More than 100 days in units of 5 days) / or ALL DATA. ・If you specify a period of data that is earlier than the period of the data saved in the RTR-51/52/53・RVR-52 Unit, all saved data will be downloaded. 3. When the downloading is completed, a message will be displayed.
● If you have set the judgment [ON] in [Set ↑↓ Limits] before downloading, the result will be displayed when the downloading is completed. Result ◆ If the amount of main unit memory is low, the downloading of recorded data may be impossible. When you check the amount of the memory to confirm whether the data you wish to download can be saved in the RTR-57C, and the data cannot be saved, the message [Memory Full] will be displayed and the downloading will be canceled.
Graph Display In this display you can view in graph for m the downloaded temperature, humidity, voltage, and pulse data. A graph is displayed for each channel of data and can be scrolled left and right with the operation dial or by the buttons on the main unit. ① CH1 ④ ⑤ ⑥ ② ③ OK ⑦ ⑧ ※⑥・⑦ are displayed when [Set ↑↓ Limits] has been set as ON. ① Displayed Channel If there is recorded data in two channels, by holding in the operation dial for more than 1.5 seconds the channel will change.
④ Data Information Displays the measurement information of the cursor bar position. Recording Date and Time Recording Interval S:Seconds M: Minutes Temperature Unit Recording Mode One Time: Endless: 【For Pulse Data】 Response Polarity Rising Edge (OFF → ON) : Falling Edge (ON → OFF) : ⑤ Cursor Bar (Fixed Display) Cursor position is displayed and measurement information is displayed at the top of the LCD display. ⑥ Upper and Lower Judge Range When [ ↑↓ Judge] is ON, the range is displayed.
Event List Display You can view the downloaded RVR-52 recorded event data on the RTR-57C in an event data list. With the operation dial or main unit buttons you can scroll up and down the data. ① ② ① Recorded Data Reading Shows the reading number of the data at the cursor position.
◆ Measurements are taken every second and events are recorded if any changes have occured. The date and time are recorded at these moments to signify an event. Mesurement Signal High Low 1 second time If both a rise and fall occur within 1 sec, the date and time is recorded at this moment.
Checking Upper and Lower Limits By setting upper and lower limits you can check data as it is downloaded to see if all measurements fall within the limit range and have the results displayed. NOTE: Upper and lower limit settings that were set by computer when setting up the recording start time for the Remote Unit will take precedence over other limit settings. ◆ Setting the Upper and Lower Limit Range 1.In the Main Menu, open [SET FUNCTIONS]-[ Set ↑↓ Limits] . 2.
5. When settings are completed, click on [Back] to activate and finish. ※ Settings can also be made via the computer. For details see the [Help] Menu in the software『T&D Recorder for Windows 』or the User’s Manual that accompanies it. ◆ Viewing the Results of Saved Data 1.In the Main Menu, open [DATA LIST] and select the data you wish to check. 2.By clicking on [Judge], the results will be displayed.
Remote Unit Conditions Display You can check selected Remote Unit info (Recording Conditions, Current Measurements, and Battery Life, etc.) ※ Before selecting a Remote Unit you can run a search to find all Remote Units with which communication is possible.Those units with which communication is possible will appear with a mark next to them. (See p.45 for details) 1.In the Main Menu, open [ Gather Data] . 2.Select the Remote Unit you wish to view. ① Select the group in which the unit is registered. 3.
【For RVR-52 Pulse Recording】 ● The total number of pulse readings during the recording interval set when recording settings were made for the Remote Unit will be displayed. ・Rising Edge (OFF → ON) : ・Falling Edge (ON → OFF) : # of Pulses / Recording Interval 【If RVR-52 has been set for Event Recording Mode】 Recording Signal Condition If no event data has been recorded, [No Event Data] will appear. ● If upper and lower limits have been set a ▼ will appear in the lower left corner of the display.
Searching for Remote Units A mark will appear next to all Remote Unit Names with which communication is possible. This function can be useful when selecting Remote Units for downloading data or starting recording. 1.In the Main Menu, open [ SEARCH RMOTE]. 2.Select the search range from ([All Groups] or [Specify Group]) and click to start the search. ● By selecting [All Groups], the search will be for Remote Units in all groups registered in the RTR-57C unit.
【Searching for Remote Units that have been additionally registered in another RTR-57C Unit】 During step 2 at left, by holding in the operation dial, you can search for Remote Units that have been additionally registered to the same group name in another RTR-57C unit. Found Remote units will automatically appear with a Remote name as shown below.
Monitoring The RTR-57C can be set to patrol and monitor RTR-51/52/53 or RVR-52 groups and display in order the current readings for view. 1.In the Main Menu, open [ MONITOR MODE]. 2.Select the monitoring range. [All Groups] : Remote Units in all groups registered to the RTR-57C will be contacted via wireless communication, and current measurement readings for those units with which communication is possible will be displayed consecutively every 2 seconds.
NOTE: ● The shorter the monitoring interval the sooner batteries will loose their power. ● ・If there are 64 Remote Units registered to the group (24 seconds per group x Number of groups) ・If there are 250 Remote Units registered to the group (54 seconds per group x Number of groups) Moreover, the display time for each unit is about 2 seconds. ● The communication time will depend on the number of registered remote units.
Displaying Recorded Data in Graph 1.In the Main Menu, open [DATA LIST] or press the [List] button on the main unit to view the data list. 2.Select the data you wish to view from the data selection display. 3.Click [Display Graph] ● In the case of event list data, click [Event List] 3.The graph or event list will appear.
【Viewing a data list】 Data Number (No. 7 of 7) The most recently downloaded data will be on top. Device Name Downloaded date 08 / 10 ・RTR51 GROUP2 Sample Group Name / Remote Unit Name (If the data logger has been registered) ● Unregistered data loggers will appear by channel name. ● If a channel name shows [……].
Deleting Data ◆ Deleting Selected Data 1.In the Main Menu, open [Data List] or press the [List] button on the main unit. Select the data you wish to delete from the list and click. 2. By clicking [Select Delete], a permission message will appear asking if it is [OK] to delete. If yes, move arrow to and click [OK] . Info about data to be deleted 3.When completed, a message will appear. Press the operation dial to finish the deletion process.
◆ Deleting All Data 1.In the Main Menu, open [Set Functions] and [Clear MEMORY] . 2.A message asking for permission will appear, if OK, move arrow to and click [OK] . 3.When completed, a message will appear. Press the operation dial to finish the deletion process. ● Recording Conditions and other settings will not be deleted. Recording conditions can be returned to the original factory settings by going to [REC SETTINGS] – [Initialize] .
Adjusting the LCD ◆ Adjusting Contrast 1.In the Main Menu, open [SET FUNCTIONS] and [Set Contrast] ● Move the operation dial up to make the display darker and down to make it lighter. 2.After completing the setting, press the dial to finish. ◆ Using the Backlight 1.In the Main Menu, open [SET FUNCTIONS] and [Backlight] . The present setting will be displayed. 2.By moving the arrow to [ON] and pressing the dial, the setting will be completed. Upon completion the new present setting will be displayed.
Muting the Operations Buzzer 1.In the Main Menu, open [SET FUNCTIONS] and [Beep] . The present setting will be displayed. 2.By moving the arrow to [OFF] and pressing the dial, the setting will be completed. Upon completion the new present setting will be displayed. 3.Move the arrow to and click [BACK] to finish. Checking Memory 1.In the Main Menu, open [SET FUNCTIONS] and [Check MEMORY] . Number of Saved Readings 2.By pressing the operation dial you can return to the SET FUNCTIONS Display.
Changing Unit of Temperature You can change the temp unit (℃ or ゜ F) displayed on the RTR-57C main unit. 1.In the Main Menu, open [SET FUNCTIONS] and [Temp Unit] The present setting will be displayed. 2.By moving the arrow to [ ゜F] and pressing the dial, the setting will be completed. Upon completion the new present setting will be displayed. 3.Move the arrow to and click [BACK] to finish.
56
… Menus at a Glance … For details about operations please see the relevant pages. GATHER DATA Call by Group 【BACK】 Download Data (See p.25) Get Remote Info (See p.43) Call by Name 【BACK】 [ALL DATA]*1 Start Gather After downloading, graph will be displayed View Remote Unit Info *1 You can select to download all data from any number of hours or days before now. SEARCH REMOTES (See p.
MONITOR MODE…View Current Measurements (See p.47) 【BACK】 All Groups Specify Group Interval Start search for contactable remotes Call by Group Consecutive display of current measurements of contacted remotes. Consecutive display of current data of all remotes in group regardless of contactability. Select from 15 sec, 30 sec, and 1 to 60 minutes (in 1 minute units) GATHER DATA (See p.34) Start Gather Upon completetion, a graph will be displayed the current data / time of downloading.
REC SETTINGS 【BACK】 Wireless (See p.19) Call by Group Package Call by Name Direct (See p.27) REC SETTINGS 【BACK】 REC Immediate REC Programmed REC Start 【BACK】 REC START Change Start Record Standby Setting Display 【BACK】 After completing settings, click [BACK] to finish. Set Rec mode 【BACK】 One Time Endless After selecting, click [BACK] to finish. Set Interval Settings Display 【BACK】 Read Settings*2 Starts the reading of present settings.
SET FUNCTIONS 【BACK】 Upper and Lower Limit Settings (See p.41) Settings Display 【BACK】 Temperature Humidity Voltage Pulse After setting, click [BACK] to finish. Backlight (See p.53) 【BACK】 ON OFF After selecting, click [BACK] to finish. Set Clock (See p.11) Settings Display 【OK】 【Cancel】 After setting, click [BACK] to finish. Check Memory (See p.54) Display 【BACK】 Clear Memory (See p.52) Confirmation Display 【BACK】 【OK】 Beep (See p.54) 【BACK】 ON OFF Temp Unit (See p.
Specifications for RTR-57C Data Collector Compatible Models RTR-51・RTR-52・RTR-53・RVR-52・TR-51A・TR-52・TR-71S・ TR-72S Recording Capacity 16 units at full data 16,000 readings × 16 = 256,000 readings Can hold up to 250 downloading sessions Functions Downloading Data (possible to display results of upper and lower limit check after downloading)・Display Saved Data Graphs・ Display Highest and Lowest Temperature・Set Recording Start for Remote Data Loggers・Delete Stored Data (one reading・ all readings)・Monitor Curr
Accessories Included Computer Communication Cable (RS-232C:D-Sub9 pin:1.
Options ◆ AC Adaptor AD-0604 Cable length : 1.
◆ Large Capacity Battery Pack RTR-05B1 4 3 62 31 14 47 Power: Lithium Battery x 1 (LS26500) Battery Life: about 2 years and 6 months (*1) (Monitoring at 1 minute interval = about 20 months) Waterproof Capability: Splash proof Operating Temperature: -40 to 80℃ (*2) Weight: about 75g (including lithium battery) Kit Contents: Lithium Battery x 1 (LS26500) Attachment hook x 1 Rubber Packing x 1 (for back of main unit) Silica Gel Pack (drying agent) x 1 Double-sided Adhesive Tape x 1 (for fastening silica
Optional Sensors ◆ Temperatue Sensors (For TR-52/RTR-52) unit : millimeters TR-5106 Teflon-Shielded Sensor Cable Length: 45mm Thermal-Constant Time : In the air : Approx. 15 Sec. In agitated water : Approx 2 Sec TR-5101 Teflon-Shielded Sensor Cable Length: 0.6m Thermal-Constant Time : In the air : Approx 15 Sec. In agitated water: Approx 2 Sec. TR-5220 Stainless Protection Sensor Cable Length: 2.0m Thermal-Constant Time : In the air:Approx 36 Sec. In agitated water: Approx 7 Sec.
◆ Sensor Extension Cable(Temp Sensor Only)unit : millimeters TR-2C30 Cable length: 3.0m Splash Resistant Materials: ① Vinyl Coated Electrical Wire NOTE: Only one cable per sensor. When using the extension cable there will be a +0.3℃ at normal temperature and at -50℃ a gap of +0.5℃ may occur.
Input Cables (Optional) ◆ Input Cable (for RVR-52) unit:millimeters RVR-7101 Input Cable Cable Length: 1.5 m Materials: ① Clip ② Vinyl Coated Electrical Wire RVR-7102 Input Cable Cable Length: 1.5 m Materials: ① M3.5 Crimp Terminal ② Vinyl Coated Electrical Wire RVR-7103 4-20mA Probe Maximum Current Input: MAX40mA Interior Resistance: 100 Ω Output: 2V at 20mA / 0.4V at 4mA Conversion Accuracy: 0.5% Cable Length: 1.5 m 67 Materials: ① M3.
Options for TR-5/RTR-5/RVR-5 ◆ Wall Attachment TR-05K1 unit : millimeters TR-05K2 Screws x 2 Materials: ① Aluminum ② Neoprene Materials: ABS Resin ◆ Maintenance Set TR-00P1 Contents: Rubber packing x 1 Silica gel x 1 Double-sided Adhesive tape x 1 Lock Screw x 2 Silicagel Rubber Packig Double Sided Adhesive Tape Lock Screw 68
■ For product information or questions contact us at: 5652-169 Sasaga Matsumoto, Nagano Japan 399-0033 Tel: +81-263-27-2131 Fax: +81-263-26-4281 E-mail: overseas@tandd.co.jp Office Hours:Monday to Friday 9:00-12:00/13:00-17:00 (GMT +9:00 Tokyo Time) [Home Page / T&D Online] We have opened an English Homepage called "T&D Online" for your convenience. Here you can find information about our company, news, products, upcoming events, software and user’s guides downloads, as well as, other support.
Provisions for Free Repair 1. If the unit does not work properly despite the fact that the customer used it properly and in line with the User's Manual, the unit shall be repaired free of charge through the distributor which sold the unit. 2. If the customer requests free repair because of trouble within the warranty period, bring or send the unit along with the warranty to the dealer. A service charge may be added if a repairperson must be sent out to the place of use for repair. 3.
Wireless Thermo Recorder Data Collector RTR-57C Warranty period Warranty 12 months from data of purchase Customer’s name: Address: Phone No.: Date of Purchase Dealer’s name: Address: Phone No.: Object of Repair Main Unit Method of Repair Send in for Repair Free repair of the unit will be carried out according to the details laid down in this manual only if the unit has broken down under normal usage as outlined in this user’s manual and during the stated warranty period.