User's Manual
35
NOTEȇ
- By initializing, all info saved in the Remote Unit will be erased and will be returned
to as it was when it left the factory.
The Default Settings are Group Name: Group 1 / Remote Unit: Sr001 /
Communication Frequency: 0
- When initializing a currently registered Remote Unit, the Registration Info will also
be erased from the registration list.
Get Remote Unit Recording Settings (Wireless Communication)
It is possible to gather the current Remote Unit Recording Settings: Recording Mode /
Type of Data / Recording Status / Current Readings / Radio Wave Strength / Battery
Level by selecting [Get Remote Unit Recording Settings (Wireless Communication)].
Recording Mode -----------The selected recording mode will be displayed: Endless or
One Time.Endless Mode--- When the logger has reached its
maximum number of readings (16,000 readings possible /
8,000 of Temp and Humidity for RTR-53), the oldest data
reading will be overwritten and recording will continue.
One-time Mode--- Recording will automatically stop when the
logger has reached its maximum number of readings (16,000
readings possible / 8,000 of Temp and Humidity for RTR-53).
Type of Data ----------------Temperature, Humidity, Voltage, Pulse, Event can be viewed
here.
Current Readings ---------The currently recorded data is displayed here.
Radio Wave Strength -----Displays the Radio Wave Strength of wireless communication
between Units.(0 – Weak < 5 – Strong)
Battery Level ----------------The remaining battery level for the Remote Unit will be
displayed.(0 – No Battery < 5 – Full)