User's Manual

55
Specifications for RTR-500DC Data Collector
RTR-500DC Unit
Compatible Devices Remote Unit:
RTR-501/501L, RTR-502/502L, RTR-503/503L, RTR-574
Repeater : RTR-500
Recording Capacity 16 units at full storage capacity16,000 readings x16= 256,000
readings. Can hold up to 250 downloading sessions
Functions Downloading data (possible to display results of upper and
lower limit check after downloading)
Display Highest and Lowest Measurement after downloading
Display Saved Data Graphs
Set Recording Start for Remote Data Loggers
Delete stored data (one reading) / Clear Memory
Monitor Current Readings
Search for Remote Units Wave Check (5-level display of radio
wave strength for each channel (0-21ch.))
LCD Display Operation Menu, Graph Display, Low Battery Life Warning,
Calendar and Clock, Contrast Adjustment, Backlight
Power 2 AAA Alkaline Batteries (LR03)
AAA Ni-MH 1.2V batteries
Optional AC adaptor (AD-0638)
USB Bus Power
Battery Life* 96 hrs* under Continuous Operation :
Using 2 AAA Alkaline Batteries / Monitor Current Readings
60 sec. pitch / No Repeater
32 hrs* under Continuous Operation :
Using 2 AAA Alkaline Batteries / Checking the Condition of
Radio Wave Transmissions
730 Downloading Sessions for RTR-501 at full storage capacity
(calculated value)
via Wireless communication / No Repeater / No Backlight
function
* Battery life varies depending upon the type of battery, the measuring
environment, the frequency of communication, and the ambient
temperature in which it is used.
Data Backup About 1 month (Saved data will be erased if all battery power is
lost.)
Communication
Interface (with PC)
USB
RS-232C(RS-232C)
Communication
Interface
(with Data Logger)
Short Range Radio Communication
Optical Communication (RTR-50x) : 2,400 to 19,200 bps
RS-232C (RTR-574) : 9,600 to 19,200 bps