Radio Wave Monitor:
Checking the Condition of Radio Wave Transmissions
Depending on the location of the logger, at special frequencies there
may be some instances in which, due to radio wave interference or
other causes, communication capability does not meet expected
standards. To help avoid interference and reduce the chances of
poor transmission, we suggest that before registering a Remote Unit
you check the condition of radio wave transmissions and when
creating a new Group assign the units to an empty frequency
channel.
The radio wave strength is displayed in a scale from 1 to 5, with the higher the number the
stronger the radio waves are.
Longitudinal axis:
Radio Wave Strength
(1 to 5)
Horizontal axis:
Frequency Channel
1.
From the Main Menu, select [SET FUNCTIONS].
2.
From the [SET FUNCTIONS] menu, select Radio Wave Monitor. (Wave
Check)
3.
Close the Radio Wave Monitor by pressing in on the operation dial.
(Returns to the [SET FUNCTIONS] Menu)
Note
- While using the Radio Wave Monitoring Function, the Auto Power Off Function will
be disabled.