6. Graph Display Range
From the entire range of recorded data this line shows the range of data
currently on display. You can use the<
List> button on the Unit to zoom in and
the <
Collect> button to zoom out on data.
7. Date and Time of Cursor Position
The date and time of the current cursor position are shown here.
Upper and Lower Limit Judgment
Upper and/or Lower Limits are set when making Recording Start settings
from a PC. When downloading recorded data from Remote Units for
which Upper / Lower Limits have been set, the judgment result will be
displayed automatically upon completion of the downloading and at the
time of graph display.
Upon Completion of
Downloading
At Time of Graph Display
(shows for about 5 seconds)
Reading Judgment Results
Starting from CH1, the results will be displayed in rotation.
:
No measurements were over the set limits.
:
One or more measurements exceeded the Upper Limit.
:
One or more measurements exceeded the Lower Limit.
:
Measurements exceeded both the Upper and Lower Limits.
:
Upper and Lower Limit Settings were not made.