Non-Wireless Communication Functions
Gather Data
Connect the RTR-500DC to the Data Logger either by placing it on
the Collector for optical communication or connect via the optional
communication cable.
Record Settings
Start (Recording Start)
Choose from Immediate Start / Programmed Start
Interval (Recording Interval)
Choose from :
10~59 S (seconds) per second / 1~ 60 M (minutes) per minute
Start Date and Time
If you are making changes to a programmed start, make sure to set start
time.
Recording Interval and Recording Mode Icons
It is not possible to set it with the Unit. The content set with the dedicated
s
oftware is displayed.
One Time: Recording will stop when the data capacity becomes full storage
Endless: Recording will continue with oldest data continuously
overwritten.