User Manual
Error Messages
MESSAGE WHAT IT MEANS WHAT TO DO
Appears when the battery
cannot provide enough
power for a test.
Rechargethebattery
immediately.
Appears when a used test
strip is inserted.
Repeattestwithanewtest
strip.Iftheerrormessage
re‐appears,pleasecontact
yourlocalcustomerservice
agentforhelp.
The temperature of the
environment, meter, or test
strip is outside the system’s
operating range. You
cannot perform a test until
the meter and test strip are
within the operating range
of 10°C to 40°C (50°F to
104°F).
Repeat the test after the
meter and test strip are
within the operating
temperatures.
Problem with the meter.
Review the instructions
and try again with a new
test strip. If the problem
persists, please contact
your local customer service
agent for help.
You may have removed the
strip after applying blood to
the absorbent hole.
Keep the test strip in the
meter after it has absorbed
the blood until the meter
displays the test result.