User Manual
2. Press Main Button to mark this test as a control solution test
With “QC” displayed, the meter will not store your test result in memory. If you
press Main Button again, the “QC” will disappear and this test is no longer a
control solution test.
WARNING:
When doing the control solution test, you have to mark it so that the test result
will not be stored in the memory. Failure to do so will mix up the blood glucose
test results with the control solution test results in memory.
3. Apply Control Solution
Shake the control solution vial thoroughly before use.
Squeeze out a drop and wipe it off, then squeeze out another
drop and place it on the tip of the vial cap. Hold the meter to
move the absorbent hole of the test strip to touch the drop. Once
the confirmation window fills completely, the meter will begin
counting down. To avoid contaminating the control solution, do
not directly apply control solution onto a strip.
4. Read and Compare the Result
After counting down to 0, the control solution test result will appear on the
display. Compare this result with the range printed on the test strip vial and it
should fall within this range. If not, please read the instructions again and
repeat the control solution test.
(100 mg/dL = 5.6 mmol/L; 109-165 mg/dL = 6.1-9.2 mmol/L)