User's Manual

A1-55
TROUBLESHOOTING
If you follow the recommended action but the problem persists, or error
messages other than the ones below appear, please call your local customer
service. Do not attempt to repair by yourself.
MESSAGE CAUSE WHAT TO DO
Inflation or pressure error
Deflation error
Please contact local customer
service for help.
Cuff pressure is too high.
No pulse rate detected or
the pulse rate is too low.
Refit cuff tightly and correctly. Relax
and the repeat measurement. If error
still remains, Please contact local
customer service for help.
Strip has been used.
Repeat the test with a new strip.
Blood sample insufficient.
Review the instructions and repeat
the test with a new strip.
Environmental
temperature is out of the
system operational range.
System operational range is 10°C -
40°C (50°F -104°F). Repeat the test
after the monitor and test strip have
reached the above temperature.
55